13.2 Input Specifications/Output Specifications/Measurement Specifications
206
(20) Active energy and reactive energy (WP+, WP-/WQLAG, WQLEAD)
Measurement method
Measured every 10 cycles (50 Hz) or 12 cycles (60 Hz) (approx. 200 ms).
During 400 Hz measurement, measured every 80 cycles using the 8-cycle waveform (approx. 200 ms).
Integrated separately by consumption and regeneration from active power.
Integrated separately by lag and lead from reactive power.
Recorded at the specified TIME PLOT interval.
Data is updated every 10 cycles (50 Hz), 12 cycles (60 Hz), or 80 cycles (400 Hz) (approx. 200 ms).
Integration starts at the same time as recording and continues to previous TIME PLOT update at
termination of recording.
Displayed item
Active energy: WP+ (consumption), WP- (regeneration)
Sum of multiple channels (for more information, see "13.8 Calculation Formula" (p.232))
Reactive energy: WQLAG (lag), WQLEAD (lead)
Sum of multiple channels (for more information, see "13.8 Calculation Formula" (p.232))
Elapsed time
Measurement range
Combination of voltage × current range (See "13.9 Range Breakdown and Combination Accuracy"
(p.245))
Value display: 6 digits
Measurement accuracy
Active energy
: Active power measurement accuracy
±
10 dgt.
Reactive energy
: Reactive power measurement accuracy
±
10 dgt.
Cumulative time accuracy : ±10 ppm ±1 s (23
°
C [73
°
F])
Event threshold
N/A
(21) Apparent power (S)
Measurement method
Calculated from RMS voltage U and RMS current I.
No polarity
Displayed item
Apparent power of each channel and its sum for multiple channels
(For details, see "13.8 Calculation Formula" (p.232))
Measurement range
Depends on the voltage × current range combination
(See "13.9 Range Breakdown and Combination Accuracy" (p.245))
Measurement accuracy
±
1 dgt. for calculations derived from the various measurement values (sum is
±
3 dgt.)
Event threshold
Power range
Event IN
Start of approx. 200 ms aggregation in which the absolute value was greater than the threshold
Event OUT
Start of approx. 200 ms aggregation in which the reading was less than (threshold - hysteresis)
following the EVENT IN state
Multiple-phase system treatment
Separate by channel
Saved waveforms
Event waveforms
HIOKI PQ3198A961-03
Содержание PQ3198
Страница 2: ...HIOKI PQ3198A961 03...
Страница 6: ...Contents iv HIOKI PQ3198A961 03...
Страница 20: ...Usage Notes 14 HIOKI PQ3198A961 03...
Страница 44: ...2 3 Display Items and Screen Types 38 HIOKI PQ3198A961 03...
Страница 58: ...3 8 Turning the Power On and Off Setting the Default Language 52 HIOKI PQ3198A961 03...
Страница 102: ...5 8 Factory Settings 96 HIOKI PQ3198A961 03...
Страница 116: ...6 4 Displaying Harmonics 110 Holding the display Hold Measured values and waveforms will be held HIOKI PQ3198A961 03...
Страница 198: ...12 4 Downloading Recorded Data to Computer 192 HIOKI PQ3198A961 03...
Страница 264: ...14 4 Disposing of the Instrument 258 HIOKI PQ3198A961 03...
Страница 296: ...A32 HIOKI PQ3198A961 03...
Страница 299: ...Index Index 3 Index Wiring 73 Z Zero adjustment 53 Zero phase A28 HIOKI PQ3198A961 03...
Страница 300: ...Index Index 4 HIOKI PQ3198A961 03...
Страница 301: ...HIOKI PQ3198A961 03...
Страница 302: ...HIOKI PQ3198A961 03...
Страница 303: ...HIOKI PQ3198A961 03...
Страница 304: ...www hioki com HIOKI PQ3198A961 03...