13.2 Input Specifications/Output Specifications/Measurement Specifications
213
10
13
Cha
p
ter 13 Spe
cificatio
ns
(32) Total harmonic voltage and Total harmonic current distortion factor (Uthd, Ithd)
Measurement method
IEC61000-4-7:2009 compliant.
Max. order: 50th
The string "-----" will appear for the voltage distortion factor when the RMS voltage is zero; for the
current distortion factor when the RMS current is zero.
Analysis window width
10 cycles (50 Hz), 12 cycles (60 Hz), or 80 cycles (400 Hz)
Number of window points
Rectangular, 4096 points
Displayed item
THD-F (total harmonic distortion factor for the fundamental wave)
THD-R (total harmonic distortion factor for the total harmonic including the fundamental wave)
Measurement range
0.00% to 100.00% (Voltage), 0.00% to 500.00% (Current)
Measurement accuracy
0.5%
Defined for the following input with a nominal input voltage of 100 V to 440 V:
Voltage, 1st order: 100% of the nominal input voltage; 5th and 7th orders: 1% of the nominal input
voltage
Current, 1st order: 100% of current range; 5th and 7th orders: 1% of current range
Event threshold
0.00% to 100.00%
Event IN
Start of approx. 200 ms aggregation in which the absolute value was greater than the threshold
Event OUT
Start of approx. 200 ms aggregation in which the reading was less than (threshold - hysteresis)
following the EVENT IN state
Multiple-phase system treatment
Separate by channel
Saved waveforms
Event waveforms
(33) K Factor (multiplication factor) (KF)
Measurement method
Calculated using the harmonic RMS current of the 2nd to 50th orders.
(For details, see "13.8 Calculation Formula" (p.232).)
Analysis window width
10 cycles (50 Hz), 12 cycles (60 Hz), or 80 cycles (400 Hz)
Number of window points
Rectangular, 4096 points
Displayed item
K factor
Measurement range
0.00 to 500.00
Measurement accuracy
-
Event threshold
0 to 500.0
Event IN
Start of approx. 200 ms aggregation in which the absolute value was greater than the threshold
Event OUT
Start of approx. 200 ms aggregation in which the reading was less than (threshold - hysteresis)
following the EVENT IN state
Multiple-phase system treatment
Separate by channel
Saved waveforms
Event waveforms
HIOKI PQ3198A961-03
Содержание PQ3198
Страница 2: ...HIOKI PQ3198A961 03...
Страница 6: ...Contents iv HIOKI PQ3198A961 03...
Страница 20: ...Usage Notes 14 HIOKI PQ3198A961 03...
Страница 44: ...2 3 Display Items and Screen Types 38 HIOKI PQ3198A961 03...
Страница 58: ...3 8 Turning the Power On and Off Setting the Default Language 52 HIOKI PQ3198A961 03...
Страница 102: ...5 8 Factory Settings 96 HIOKI PQ3198A961 03...
Страница 116: ...6 4 Displaying Harmonics 110 Holding the display Hold Measured values and waveforms will be held HIOKI PQ3198A961 03...
Страница 198: ...12 4 Downloading Recorded Data to Computer 192 HIOKI PQ3198A961 03...
Страница 264: ...14 4 Disposing of the Instrument 258 HIOKI PQ3198A961 03...
Страница 296: ...A32 HIOKI PQ3198A961 03...
Страница 299: ...Index Index 3 Index Wiring 73 Z Zero adjustment 53 Zero phase A28 HIOKI PQ3198A961 03...
Страница 300: ...Index Index 4 HIOKI PQ3198A961 03...
Страница 301: ...HIOKI PQ3198A961 03...
Страница 302: ...HIOKI PQ3198A961 03...
Страница 303: ...HIOKI PQ3198A961 03...
Страница 304: ...www hioki com HIOKI PQ3198A961 03...