The accuracy of active power calculations can be calculated as follows, taking into account the phase
accuracy:
Example measurement conditions
Wiring: 3-phase/4-wire (3P4W)
Current sensor: Model CT7136
Current range: 50 A (power range: 90 kW)
"13.9 Range Breakdown and Combination Accuracy" (p.245)
Measured values: Active power of 30 kW, power factor lag 0.8
Accuracy
Active power accuracy for current sensor combination (Model CT7136 sensor, 50 A range):
±0.5% rdg.±0.22% f.s.
Internal circuit voltage of the instrument - current phase difference: ±0.2865°
(Effect of power factor: 1.0% rdg. or less)
Phase accuracy of the CT7136: ±0.5°
"13.2 Input Specifications/Output Specifications/Measurement Specifications" (p.194)
"13.9 Range Breakdown and Combination Accuracy" (p.245)
Phase accuracy shown in “Specifications” of the CT7136 Instruction Manual
Power factor accuracy based on phase accuracy
Phase accuracy (in combination with current sensor) = Instrument internal circuit phase accuracy
(±0.2865°) + CT7136 phase accuracy (±0.5°) = ±0.7865°
Phase difference
θ
= cos
−
1
(power factor)= cos
−
1
0.8= 36.87°
Power factor error range based on phase accuracy = cos(36.87°±0.7865°) = 0.7916 to 0.8082
Power factor accuracy based on phase accuracy
Active power accuracy = current sensor combined ac power factor accuracy based on phase
accuracy
= ±0.5% rdg. ±0.22% f.s. ±1.05% rdg.
= ±1.55% rdg.±0.22% f.s.
Accuracy relative to measured values = active power 30 kW × ± 1.55% rdg. + 90 kW range × 0.22% f.s.
= ±0.663 kW
= ±0.663 kW/30 kW = ±2.21% rdg.
Appendix 8 Method for Calculating Active Power
Accuracy
Power factor accuracy based on phase accuracy (minimum) =
% = -1.05%
Power factor accuracy based on phase accuracy (maximum) =
% = +1.025%
Power factor accuracy based on phase accuracy:
±1.05% rdg.
0.7916 0.8
–
0.8
------------------------------ 100
×
0.8082 0.8
–
0.8
------------------------------ 100
×
The value, whichever is
worse, is specified to be
the phase accuracy.
HIOKI PQ3198A961-03
Содержание PQ3198
Страница 2: ...HIOKI PQ3198A961 03...
Страница 6: ...Contents iv HIOKI PQ3198A961 03...
Страница 20: ...Usage Notes 14 HIOKI PQ3198A961 03...
Страница 44: ...2 3 Display Items and Screen Types 38 HIOKI PQ3198A961 03...
Страница 58: ...3 8 Turning the Power On and Off Setting the Default Language 52 HIOKI PQ3198A961 03...
Страница 102: ...5 8 Factory Settings 96 HIOKI PQ3198A961 03...
Страница 116: ...6 4 Displaying Harmonics 110 Holding the display Hold Measured values and waveforms will be held HIOKI PQ3198A961 03...
Страница 198: ...12 4 Downloading Recorded Data to Computer 192 HIOKI PQ3198A961 03...
Страница 264: ...14 4 Disposing of the Instrument 258 HIOKI PQ3198A961 03...
Страница 296: ...A32 HIOKI PQ3198A961 03...
Страница 299: ...Index Index 3 Index Wiring 73 Z Zero adjustment 53 Zero phase A28 HIOKI PQ3198A961 03...
Страница 300: ...Index Index 4 HIOKI PQ3198A961 03...
Страница 301: ...HIOKI PQ3198A961 03...
Страница 302: ...HIOKI PQ3198A961 03...
Страница 303: ...HIOKI PQ3198A961 03...
Страница 304: ...www hioki com HIOKI PQ3198A961 03...