13.1 General Specifications
193
10
13
Cha
p
ter 13 Spe
cificatio
ns
Specifications
Chapter 13
13.1 General Specifications
Operating environment
Indoor use, pollution degree 2, altitude up to 3000 m (9843 ft.)
At an altitude of above 2000 m (6562 ft.), the measurement categories are lowered to 600 V CAT III.
Operating temperature and humidity 0°C to 30°C (32°F to 86°F) 95% RH or less (no condensation)
When charging battery: 10°C to 30°C (50°F to 86°F)
30°C to 50°C (86°F to 122°F) 80% RH or less (no condensation)
When charging battery: 30°C to 35°C (86°F to 95°F)
Storage temperature and humidity -20°C to 30°C (-4°F to 86°F) 95% RH or less (no condensation)
30°C to 50°C (86°F to 122°F) 80% RH or less (no condensation)
(If the instrument will not be used for an extended period of time, remove the battery pack and store
in a cool location [from -20°C to 30°C (-4°F to 86°F)].)
Dust and water resistance
IP30 (EN60529)
Standards
Safety EN61010
EMC
EN61326 Class A
Power supply quality
measurement method
IEC 61000-4-30 Ed3:2015 Class A, IEEE 1159
Power supply
Z1002 AC Adapter
Rated supply voltage
: 100 V AC to 240 V AC (Voltage fluctuations of
±
10% from the rat-
ed supply voltage are taken into account.)
Rated power supply frequency : 50 Hz/60 Hz
Anticipated transient overvoltage: 2500 V
Maximum rated power
: 80 VA (When charging, including the AC adapter)
35 VA (When charging, main unit only)
Z1003 Battery Pack
Rated supply voltage
: 7.2 V DC
Maximum rated power
: 8 VA
Recharge function
Charges the battery regardless of whether the instrument is on or off.
Charging time 5 hours 30 minutes at a maximum (at 23°C, as a referential)
Continuous operating time
When Z1003 Battery Pack is used (at 23°C, as a referential)
Approx. 3 hours (fully charged, continuous operation, LCD backlight AUTO OFF)
Backup battery
Approx. 10 years (at 23°C, as a referential)
For backup clock and setting conditions (Lithium battery)
Maximum recording interval
Repeat recording function, 1 week: 55 weeks
Repeat recording function, 1 day: 366 days
Repeat recording function, off: 35 days
Maximum recording events
9999 events (Switchable between 1000 events and 9999 events)
Clock function
Auto-calendar, leap-year correcting 24-hour clock
Real-time clock accuracy
±0.3 s per day (instrument on, 23°C±5°C [73°F±9°F])
±1 s per day (instrument on, within operating temperature and humidity range)
±3 s per day (instrument off) (at 23°C, as a referential)
Display refresh rate
Approx. 0.5 s
Display
6.5-inch TFT color LCD (640 × 480 dots)
Interface
SD memory card, USB, LAN, RS-232C, External I/O
Dimensions
Approx. 300W × 211H × 68D mm (11.81
″
W × 8.31
″
H × 2.68
″
D)
Body
Strap can be attached.
Mass
Approx. 2.5 kg (88.2 oz.) (with Z1003 Battery Pack installed)
Product warranty period
3 years
Accessories
See: "Accessories" (p.4)
Options
See: "Options" (p.5)
HIOKI PQ3198A961-03
Содержание PQ3198
Страница 2: ...HIOKI PQ3198A961 03...
Страница 6: ...Contents iv HIOKI PQ3198A961 03...
Страница 20: ...Usage Notes 14 HIOKI PQ3198A961 03...
Страница 44: ...2 3 Display Items and Screen Types 38 HIOKI PQ3198A961 03...
Страница 58: ...3 8 Turning the Power On and Off Setting the Default Language 52 HIOKI PQ3198A961 03...
Страница 102: ...5 8 Factory Settings 96 HIOKI PQ3198A961 03...
Страница 116: ...6 4 Displaying Harmonics 110 Holding the display Hold Measured values and waveforms will be held HIOKI PQ3198A961 03...
Страница 198: ...12 4 Downloading Recorded Data to Computer 192 HIOKI PQ3198A961 03...
Страница 264: ...14 4 Disposing of the Instrument 258 HIOKI PQ3198A961 03...
Страница 296: ...A32 HIOKI PQ3198A961 03...
Страница 299: ...Index Index 3 Index Wiring 73 Z Zero adjustment 53 Zero phase A28 HIOKI PQ3198A961 03...
Страница 300: ...Index Index 4 HIOKI PQ3198A961 03...
Страница 301: ...HIOKI PQ3198A961 03...
Страница 302: ...HIOKI PQ3198A961 03...
Страница 303: ...HIOKI PQ3198A961 03...
Страница 304: ...www hioki com HIOKI PQ3198A961 03...