7.5 Displaying Flicker Values in Graph and List Form
137
Chap
ter 7
Monitoring
Flu
ctuations
i
n Meas
ure
d
V
alues
(T
IME PLOT Screen)
7
This section describes how to display a
Δ
V10 flicker fluctuation graph.
Δ
V10 flicker reference voltage
In
Δ
V10 flicker measurement, the reference voltage is automatically set internally using AGC (automatic
gain control).
Once the fluctuating voltage value has stabilized, the reference voltage is automatically changed to that
value. Consequently, there is no need to switch supply voltage settings as with conventional
Δ
V10 flicker
meters.
Example:
Fluctuating voltage: Stabilizes at 96 V rms The reference voltage is automatically changed to 96 V rms.
Fluctuating voltage: Stabilizes at 102 V rms The reference voltage is automatically changed to 102 V rms.
Due to the influence of the high pass filter used in
Δ
V10 flicker measurement, measured values are
unstable when starting
Δ
V10 measurement immediately after settings have been configured, and the first
and second
Δ
V10 measured values may be excessively high. It is recommended to wait about 5 minutes
after making settings on the
[SYSTEM]
screen before starting measurement.
Displaying a
Δ
V10 flicker fluctuation graph
[TIME PLOT]
screen
[Flicker]
[Graph]
Select with the
F
key.
To enlarge or reduce the graph (p.138)
To read the value above the cursor (p.139)
To scroll waveform data (p.139)
1
• The graph is updated once a minute, regardless of the TIME PLOT interval set in
[SYS-
TEM]
-
DF1
[Record]
-
F1
[
Interval
]
.
• The graph is not displayed unless
[Flicker]
is set to
[
Δ
V 10]
in
[SYSTEM]
-
DF1
[Main]
-
F2
[Measure2]
.
•
Δ
V10 flicker can be measured simultaneously for the voltage channels U1, U2, and U3
(depends on connection).
HIOKI PQ3198A961-03
Содержание PQ3198
Страница 2: ...HIOKI PQ3198A961 03...
Страница 6: ...Contents iv HIOKI PQ3198A961 03...
Страница 20: ...Usage Notes 14 HIOKI PQ3198A961 03...
Страница 44: ...2 3 Display Items and Screen Types 38 HIOKI PQ3198A961 03...
Страница 58: ...3 8 Turning the Power On and Off Setting the Default Language 52 HIOKI PQ3198A961 03...
Страница 102: ...5 8 Factory Settings 96 HIOKI PQ3198A961 03...
Страница 116: ...6 4 Displaying Harmonics 110 Holding the display Hold Measured values and waveforms will be held HIOKI PQ3198A961 03...
Страница 198: ...12 4 Downloading Recorded Data to Computer 192 HIOKI PQ3198A961 03...
Страница 264: ...14 4 Disposing of the Instrument 258 HIOKI PQ3198A961 03...
Страница 296: ...A32 HIOKI PQ3198A961 03...
Страница 299: ...Index Index 3 Index Wiring 73 Z Zero adjustment 53 Zero phase A28 HIOKI PQ3198A961 03...
Страница 300: ...Index Index 4 HIOKI PQ3198A961 03...
Страница 301: ...HIOKI PQ3198A961 03...
Страница 302: ...HIOKI PQ3198A961 03...
Страница 303: ...HIOKI PQ3198A961 03...
Страница 304: ...www hioki com HIOKI PQ3198A961 03...