13.3 Screen Specifications
220
(5) Event Settings
Event hysteresis
0% to 10% (Applies to all parameters except frequency.)
Fixed to 0.1 Hz for frequency; percentage of threshold value for other parameters.
Maximum recordable events
1000/9999
Sets the maximum number of recordable events per measurement when repeat recording is off.
When the repeat recording function is on, the number of events is obtained by multiplying this set-
ting by the repeat count. A setting of 9999 disables voltage waveform comparison events.
Slide reference voltage
OFF/ON (Applies to swells and dips.)
When turned on, the slide reference voltage is used instead of the nominal voltage.
Timer event count
OFF/1 minute/5 minutes/10 minutes/30 minutes/1 hour/2 hours
Events are generated at the chosen interval.
Continuous event count
OFF/1/2/3/4/5 times
Applies to all events being recorded. When time target events occur, they are automatically treated
as sequential events if the event in question occurs the set number of times. However, events oc-
curring during sequential events cannot trigger sequential events. In addition, generation of se-
quential events stops when recording stops.
External event
OFF,ON
Event setting details
See: "5.6 Changing Event Settings" (p.87)
(6) [MEMORY] Screen
Target interface
SD memory card
Function
Mass storage, saving (of settings data), loading (of settings data, measurement data, event data,
screen data, and version upgrade files), deletion of folders and files, and formatting
HIOKI PQ3198A961-03
Содержание PQ3198
Страница 2: ...HIOKI PQ3198A961 03...
Страница 6: ...Contents iv HIOKI PQ3198A961 03...
Страница 20: ...Usage Notes 14 HIOKI PQ3198A961 03...
Страница 44: ...2 3 Display Items and Screen Types 38 HIOKI PQ3198A961 03...
Страница 58: ...3 8 Turning the Power On and Off Setting the Default Language 52 HIOKI PQ3198A961 03...
Страница 102: ...5 8 Factory Settings 96 HIOKI PQ3198A961 03...
Страница 116: ...6 4 Displaying Harmonics 110 Holding the display Hold Measured values and waveforms will be held HIOKI PQ3198A961 03...
Страница 198: ...12 4 Downloading Recorded Data to Computer 192 HIOKI PQ3198A961 03...
Страница 264: ...14 4 Disposing of the Instrument 258 HIOKI PQ3198A961 03...
Страница 296: ...A32 HIOKI PQ3198A961 03...
Страница 299: ...Index Index 3 Index Wiring 73 Z Zero adjustment 53 Zero phase A28 HIOKI PQ3198A961 03...
Страница 300: ...Index Index 4 HIOKI PQ3198A961 03...
Страница 301: ...HIOKI PQ3198A961 03...
Страница 302: ...HIOKI PQ3198A961 03...
Страница 303: ...HIOKI PQ3198A961 03...
Страница 304: ...www hioki com HIOKI PQ3198A961 03...