14.2 Trouble Shooting
252
Verify below before returning the instrument for repair.
Perform a system reset.
This will return all settings to their factory defaults.
See:
"5.7 Initializing the Instrument (System Reset)" (p.94)
Before having the instrument repaired
Symptom
Check item or cause
Remedy and reference
The display does
not appear when
you turn the power
on.
Has the power cord been disconnected?
Is it connected properly?
Verify that the power cord is connected
properly.
See: "3.4 Connecting the AC Adapter" (p.45)
Keys do not work.
Has the key lock been activated?
Press and hold the
ESC
key for at least 3
seconds to cancel the key lock. If you set a
passcode, enter the same passcode to dis-
engage the key lock.
See:
"7 Engage the key lock." (p.28)
Voltage or current
measured values
are not being dis-
played.
Are the voltage cords or current sensors con-
nected improperly?
Verify connections.
See: "3.6 Connecting the Voltage Cords" (p.47)
to "4.6 Verifying Correct Wiring (Connec-
tion Check)" (p.66)
Are the input channels and display channels
incorrect?
-
The instrument can-
not measure the fre-
quency.
Measured values do
not stabilize.
Is the input frequency within the guaranteed
accuracy range?
For a measurement frequency of 50 Hz, 40 Hz to
58 Hz.
For a measurement frequency of 60 Hz, 51 Hz to
70 Hz.
For a measurement frequency of 400 Hz, 360 Hz
to 440 Hz.
Measurement cannot be performed if the input
frequency is outside the guaranteed accuracy
fundamental wave range.
-
Is the input frequency lower than the setting?
Is a signal being input to U1?
Stable measurement may not be possible if input
of at least 2% f.s. is not being supplied to U1 (the
reference channel).
When no apparent cause can be established
HIOKI PQ3198A961-03
Содержание PQ3198
Страница 2: ...HIOKI PQ3198A961 03...
Страница 6: ...Contents iv HIOKI PQ3198A961 03...
Страница 20: ...Usage Notes 14 HIOKI PQ3198A961 03...
Страница 44: ...2 3 Display Items and Screen Types 38 HIOKI PQ3198A961 03...
Страница 58: ...3 8 Turning the Power On and Off Setting the Default Language 52 HIOKI PQ3198A961 03...
Страница 102: ...5 8 Factory Settings 96 HIOKI PQ3198A961 03...
Страница 116: ...6 4 Displaying Harmonics 110 Holding the display Hold Measured values and waveforms will be held HIOKI PQ3198A961 03...
Страница 198: ...12 4 Downloading Recorded Data to Computer 192 HIOKI PQ3198A961 03...
Страница 264: ...14 4 Disposing of the Instrument 258 HIOKI PQ3198A961 03...
Страница 296: ...A32 HIOKI PQ3198A961 03...
Страница 299: ...Index Index 3 Index Wiring 73 Z Zero adjustment 53 Zero phase A28 HIOKI PQ3198A961 03...
Страница 300: ...Index Index 4 HIOKI PQ3198A961 03...
Страница 301: ...HIOKI PQ3198A961 03...
Страница 302: ...HIOKI PQ3198A961 03...
Страница 303: ...HIOKI PQ3198A961 03...
Страница 304: ...www hioki com HIOKI PQ3198A961 03...