6.3 Displaying Phase Relationships ([VECTOR] Screen)
103
Chapte
r 6 Mon
itoring Insta
ntaneou
s V
alues
(VIEW
Screen)
6
Axis display
You can select whether to use a linear display (LINEAR) or logarithmic display (LOG) for the vector axis.
If you select the logarithmic display method, the vector is easy to see even at low levels.
RMS/phase angle/content percentage display
Selects which value to display (RMS value display, phase angle display, or content percentage display). If
[Phase]
is selected, you can also set the phase angle value display method.
Phase angle value display method
You can select the type of phase angle display. (This setting can be configured only when
[Phase]
is
selected.)
Setting this parameter to
[lag360]
allows the display to be rotated clockwise 0° to 360°.
If
[lag360]
is selected, you can also set the phase angle reference source.
Phase angle reference source
You can select the reference source (0°) to display the phase angle
value.
Setting Contents:(
∗
: Default setting)
LINEAR
∗
Linear display
LOG
Log Logarithmic display
When the 400 Hz measurement frequency is selected, harmonic analysis is performed up
to the 10th order, and inter-harmonic analysis is not available.
Setting Contents:(
∗
: Default setting)
Level
∗
RMS
Phase
Phase angle
Content
Content percentage
Setting Contents:(
∗
: Default setting)
±
180
∗
lead 0 to 180°, lag 0 to -180°
lag360
lag 0 to 360°
U1
∗
Uses U1 as the reference source.
I1
Uses I1 as the reference source.
U2
Uses U2 as the reference source.
I2
Uses I2 as the reference source.
U3
Uses U3 as the reference source.
I3
Uses I3 as the reference source.
HIOKI PQ3198A961-03
Содержание PQ3198
Страница 2: ...HIOKI PQ3198A961 03...
Страница 6: ...Contents iv HIOKI PQ3198A961 03...
Страница 20: ...Usage Notes 14 HIOKI PQ3198A961 03...
Страница 44: ...2 3 Display Items and Screen Types 38 HIOKI PQ3198A961 03...
Страница 58: ...3 8 Turning the Power On and Off Setting the Default Language 52 HIOKI PQ3198A961 03...
Страница 102: ...5 8 Factory Settings 96 HIOKI PQ3198A961 03...
Страница 116: ...6 4 Displaying Harmonics 110 Holding the display Hold Measured values and waveforms will be held HIOKI PQ3198A961 03...
Страница 198: ...12 4 Downloading Recorded Data to Computer 192 HIOKI PQ3198A961 03...
Страница 264: ...14 4 Disposing of the Instrument 258 HIOKI PQ3198A961 03...
Страница 296: ...A32 HIOKI PQ3198A961 03...
Страница 299: ...Index Index 3 Index Wiring 73 Z Zero adjustment 53 Zero phase A28 HIOKI PQ3198A961 03...
Страница 300: ...Index Index 4 HIOKI PQ3198A961 03...
Страница 301: ...HIOKI PQ3198A961 03...
Страница 302: ...HIOKI PQ3198A961 03...
Страница 303: ...HIOKI PQ3198A961 03...
Страница 304: ...www hioki com HIOKI PQ3198A961 03...