76
_____________________________________________________________________________________________
5.4 Commands Specific to the 3532-50, 3522-50
_____________________________________________________________________________________________
:SPEEd
Sets the testing speed.
:SPEEd?
Queries the testing speed.
Syntax
:SPEEd
<data>
<data>
FAST/NORMal/SLOW/SLOW2 (character data)
Function
Sets the testing speed.
Example
Transmission
:SPEEd NORMal
Error
If <data> is other than character data described above, an execution error occurs.
Executing this command while the open or short circuit compensation is
performed generates an execution error.
Syntax
:SPEEd?
Function
Returns the setting of testing speed as character data.
FAST, NORMAL, SLOW, SLOW2
Example
Response
If headers are on
:SPEED NORMAL
If headers are off
NORMAL
Error
If the response message is longer than 300 bytes, a query error is generated.