77
_____________________________________________________________________________________________
5.4 Commands Specific to the 3532-50, 3522-50
_____________________________________________________________________________________________
:TRIGger
Sets the type of trigger.
:TRIGger?
Queries the trigger setting.
Syntax
:TRIGger
<data>
<data>
INTernal/EXTernal (character data)
Function
Sets the type of trigger.
INTernal Internal trigger mode
EXTernal External trigger mode
Example
Transmission
:TRIGger INTernal
The trigger mode is set to internal trigger.
Error
If <data> is other than character data described above, an execution error occurs.
Executing this command while the open or short circuit compensation is
performed generates an execution error.
Syntax
:TRIGger?
Function
Returns the trigger setting as character data.
INTERNAL/EXTERNAL
Example
Response
If headers are on
:TRIGGER INTERNAL
If headers are off
INTERNAL
The trigger mode has been set to internal triggering.
Error