:LEVel:CVOLTage
Sets the constant voltage value.
:LEVel:CVOLTage?
Queries the constant voltage values.
Syntax
:LEVel:CVOLTage
<data>
<data>
Numerical data in NR3 format
3532-50: 0.010 to 5.000 (frequency 42 Hz to 1 MHz)
0.010 to 1.000 (frequency 1.001 MHz to 5 MHz)
3522-50: 0.010 to 5.000
Function
Sets the value of the constant voltage.
The numerical value can be in NRf format, but rounding is performed for figures
beyond the last valid decimal place.
Example
Transmission
:LEVel:CVOLTage 1.234
The constant voltage value is set to 1.234 V.
Error
If <data> is other than numerical value described above, an execution error
occurs.
Executing this command while the open or short circuit compensation is
performed generates an execution error.
Syntax
:LEVel:CVOLTage?
Function
Returns the constant voltage value as a numerical value in NR2 format.
Example
Response
If headers are on
:LEVEL:CVOLTAGE 1.234
If headers are off
1.234
The constant voltage level has been set to 1.234 V.
Error