60
_____________________________________________________________________________________________
5.4 Commands Specific to the 3532-50, 3522-50
_____________________________________________________________________________________________
:LIMiter:CURRent
Sets the current limit value.
:LIMiter:CURRent?
Queries the current limit value.
Syntax
:LIMiter:CURRent
<data>
<data>
Numerical data in NR3 format from 0.01E-03 to 99.99E-03
Function
Sets the current limit value.
The numerical value can be in NRf format, but rounding is performed for figures
beyond the last valid decimal place.
Example
Transmission
:LIMiter:CURRent 10.00E-03
The current limit value is set to 10 mA.
Error
If <data> is other than numerical value described above, an execution error
occurs.
Executing this command while the open or short circuit compensation is
performed generates an execution error.
Syntax
:LIMiter:CURRent?
Function
Returns the current limit value as a numerical value in NR3 format.
Example
Response
If headers are on
:LIMITER:CURRENT 10.00E-03
If headers are off
10.00E-03
The current limit value has been set to 10 mA.
Error
If the response message is longer than 300 bytes, a query error is generated.