49
_____________________________________________________________________________________________
5.4 Commands Specific to the 3532-50, 3522-50
_____________________________________________________________________________________________
:CORRection:SHORt
Enables and disables the short circuit compensation function.
Syntax
:
CORRection:SHORt
<data>
<data>
OFF/ALL (character data) or numerical data in NR3 format
3532-50: 42.0E+00 to 5.000E+06
3522-50: 0 to 100.0E+03
Function
Enables and disables the short circuit compensation function.
The numerical value can be in NRf format, but rounding is performed for figures
beyond the last valid decimal place.
OFF
The short circuit compensation is not performed.
ALL
The short circuit compensation is performed at all the test
frequencies.
Numerical data The short circuit compensation is performed at the set test
frequency (spot compensation).
0: compensation for DC
Note
When the compensation is performed at all the test frequencies, about 3 minutes
compensation (using the 3532-50) or about 2 minutes compensation (using the
3522-50) is required. Executing the commands which changes test settings
during compensation for all frequency generate an execution error. Be sure not to
execute commands other than that of checking each status registers such as
*
ESR? and :ESR0?.
When the SPOT compensation is performed, it takes about maximum 15 minutes
(1 mHz compensation) for the 3522-50 to read the compensation data.
Example
Transmission
:CORRection:SHORt 1E+3
The short circuit compensation function at 1 kHz is enabled.
Error
If <data> is other than character data or numerical value described above, an
execution error occurs.
Executing this command while the comparator function is performed generates an
execution error.