5.4 Commands Specific to the 3532-50, 3522-50
_____________________________________________________________________________________________
:CORRection:SHORt?
Queries the short circuit compensation function enablement.
:DISPlay:MONItor?
Queries the voltage and current monitored parameters.
Syntax
:
CORRection:SHORt?
Function
Returns the current setting of the short circuit compensation enablement as
character data or a numerical value in NR3 format.
OFF
The short circuit compensation function has been set to off.
ALL
The short circuit compensation function at all the test
frequencies has been set to on.
Numerical data The short circuit compensation function at the set test
frequency has been set to on (spot compensation).
Example
Response
If headers are on :
CORRECTION:SHORT 1.000E+03
If headers are off
1.000E+03
The open circuit compensation function at 1 kHz has been enabled.
Error
If the response message is longer than 300 bytes, a query error is generated.
Syntax
:DISPlay:MONItor?
Function
Returns the monitored parameters as <voltage monitored value> and <current
monitored value> in order.
Voltage monitored value Numerical value in NR2 format
Current monitored value
Numerical value in NR3 format
Example
Response
If headers are on
:DISPLAY:MONITOR 1.23,0.12E-03
If headers are off
1.23,0.12E-03
Error