48
_____________________________________________________________________________________________
5.4 Commands Specific to the 3532-50, 3522-50
_____________________________________________________________________________________________
:CORRection:OPEN?
Queries the open circuit compensation function enablement.
Syntax
:CORRection:OPEN?
Function
Returns the current setting of open circuit compensation function enablement as
character data or a numerical value in NR3 format.
OFF
The open circuit compensation function has been set to off.
ALL
The open circuit compensation function at all the test
frequencies has been set to on.
Numerical data The open circuit compensation function at the set test
frequency has been set to on (spot compensation).
Example
Response
If headers are on
:CORRECTION:OPEN 1.000E+03
If headers are off
1.000E+03
The open circuit compensation at 1 kHz has been enabled.
Error
If the response message is longer than 300 bytes, a query error is generated.