4.9 Output Queue
_____________________________________________________________________________________________
4.9 Output Queue
4.10 Input Buffer
Response messages accumulate in the output queue and all data are received
and cleared.
The output queue is also cleared when the power is turned off and turned on
again.
The 3532-50/3522-50 have an output queue of 300 bytes capacity. If the
response messages overflow this limit of 300 bytes, a query error is
generated, and the output buffer is cleared
The 3532-50/3522-50 have an input buffer of 300 bytes capacity.
When more than 300 bytes of data are transmitted, when the buffer is full
any subsequent bytes received will be ignored.
(When the controller handshake setting is not the same as the 9593-01.)