80
_____________________________________________________________________________________________
5.5 Response Format for Queries as Numerical Value
_____________________________________________________________________________________________
.
E
1 2
(-)
.
(-)
.
(-) .
1 2
.
Up to the fifth decimal point
.
Up to the second decimal point
.
.
5.5 Response Format for Queries as Numerical Value
(1) Test value
1. The response formats for |Z| (impedance),|Y| (admittance), Cs (static capacitance in
series equivalent circuit mode), Cp (static capacitance in parallel equivalent circuit
mode), Ls (inductance in series equivalent circuit mode), Lp (inductance in parallel
equivalent circuit mode), Rs (effective resistance in series equivalent circuit mode),
G (conductance), Rp (effective resistance in parallel equivalent circuit mode), X
(reactance), B (susceptance) are as follows. (in NR3 format)
1
Mantissa Five digits and decimal point
2
Exponent Two digits
When the value is overflow or underflow, the following value is displayed.
Overflow 99
Underflow -99
2. The response formats (in NR2 format) for
θ
(phase angle).
1
Sign Only when the value is negative, minus (-) is prefixed.
2
Numerical Up to the second decimal point
When the value is overflow or underflow, the following value is displayed.
Overflow 999.9
Underflow -999.9
3. The response formats (in NR2 format) for D (loss coefficient)
When the value is overflow or underflow, the following value is displayed.
Overflow 999999
Underflow -999999
4. The response formats (in NR2 format) for Q (Q factor)
When the value is overflow or underflow, the following value is displayed.
Overflow 9999
Underflow -9999