69
_____________________________________________________________________________________________
5.4 Commands Specific to the 3532-50, 3522-50
_____________________________________________________________________________________________
: settable /
---: cannot be set
Range
number
Range
(
Ω
)
to 100.0 kHz
100.1 kHz to
1.000 MHz
1.001 MHz to
5.000 MHz
1
0.1
2
1
3
10
4
100
5
1 k
6
10 k
7
100 k
8
1 M
---
9
10 M
---
---
10
100 M
---
---
For 3532-50 If the test frequency is greater than 100 kHz, the range
number 9 (10 M
Ω
range) cannot be set. If the test
frequency is greater than 1 MHz, the range number 8
(1 M
Ω
range) cannot be set.
:RANGe
Sets the test range.
Syntax
:RANGe
<data>
<data>
Numerical data in NR1 format
1 to 10
Function
Sets the test range.
The numerical value can be in NRf format, but any digits after the decimal point
will be rounded.
If this command is executed, the setting of the :RANGe:AUTO command is
automatically changed to OFF.
The numerical value corresponding to the test range and frequency which can be
set is as follows.
Example
Transmission
:RANGe 5
The test range is set to 1 k
Ω
.
Error
If <data> is other than numerical value described above, an execution error
occurs.
Executing this command while the open or short circuit compensation is
performed generates an execution error.