5.4 Commands Specific to the 3532-50, 3522-50
_____________________________________________________________________________________________
(for 3532-50)
Note
:FREQuency
Sets the test frequency.
:FREQuency?
Queries the test frequency.
Syntax
FREQuency
<data>
<data>
Numerical data in NR3 format
3532-50: 42.0E+00 to 5.000E+06
3522-50: 0 to 100.0E+03
Function
Sets the testing frequency.
The numerical value can be in NRf format, but rounding is performed for figures
beyond the last valid decimal place.
Specify <data> to 0 at DC measurement.
If the test frequency is greater than 100 kHz, the test range which can be set is
limit. If it is greater than 1 MHz, the range of test signal level is limit. When
the test range and the test signal level are greater than each range after changing
frequencies, they are automatically change over the highest range compatible with
this test frequency setting. For details, refer to the 3532-50 instruction manual.
Example
Transmission
:FREQuency 1.000E+03
The test frequency is set to 1 kHz.
Error
If <data> is other than numerical value described above, an execution error
occurs.
Executing this command while the open or short circuit compensation is
performed generates an execution error.
Syntax
:FREQuency?
Function
Returns the currently test frequency as a numerical value in NR3 format.
Example
Response
If headers are on
:FREQUENCY 1.000E+03
If headers are off
1.000E+03
The test frequency has been set to 1 kHz.
Error
If the response message is longer than 300 bytes, a query error is generated.