74
_____________________________________________________________________________________________
5.4 Commands Specific to the 3532-50, 3522-50
_____________________________________________________________________________________________
:SCALe:FVALue
Sets the first parameters (a and b) in the scaling function.
:SCALe:FVALue?
Queries the first parameters (a and b) in the scaling function.
Syntax
:SCALe:FVALue
<a>, <b>
<data>
Numerical data in NR3 format
Function
Sets the first parameters (a and b values) in the scaling function.
For calculation equation of the scaling function, see the Instruction Manual of
main unit.
Example
Transmission
:SCALe:FVALue 2.0000E+00,1.0000E+00
Sets a value to 2.0000, and b value to 1.0000.
Error
If <data> is other than numerical data described above, an execution error occurs.
Executing this command while the open or short circuit compensation is
performed generates an execution error
.
Syntax
:SCALe:FVALue?
Function
Returns the setting of the first parameters (a and b values) in the scaling function
as a numerical value in NR3 format.
Example
Response
If headers are on
:SCALE:FVALUE 2.0000E+00,1.0000E+00
If headers are off
2.0000E+00,1.0000E+00
Error
If the response message is longer than 300 bytes, a query error occurs.