66
────────────────────────────────────────────────────
4.8 Short Circuit Compensation
────────────────────────────────────────────────────
NOTE
NOTE
When compensation is to be performed for some new frequency:
Input the frequency for compensation using the numeric keypad.
Frequency range which can be set : 0.000 Hz (DC) to 100 kHz
If you attempt to set a test frequency greater than 100 kHz, it will
automatically be reduced to 100 kHz.
If you attempt to set a test frequency lower than 1 mHz, it will automatically
be increased to 0 Hz (DC).
If you make a mistake during input of the numerical value, press the
key to cancel the input so far, and start again.
When you have completed inputting the numerical value for the frequency,
press the appropriate units key
,
, or
so as to confirm the
displayed frequency for compensation.
Until the numerical value for the frequency has been input, the units keys
cannot be used.
3.
After the frequency for SPOT compensation has been input as above, press
the
key.
Check again - are the test leads properly shorted together with the shorting
bar?
The time taken for the SPOT compensation varies depending on the
frequency for compensation, but it takes about 15 minutes maximum at 1
mHz compensation.