90
────────────────────────────────────────────────────
4.14 Setting and Activating the Comparator
────────────────────────────────────────────────────
4.14.6 Setting the Upper or Lower Limit Value as a Percentage
(%) Relative to a Standard Value
Condition setting
screen
NOTE
REF key
HI key
LO key
Percentage Value Setting Mode
Numeric
keypad
Units keys
1.
Press this key on the Conditions setting screen to select the percentage value
setting mode.
At this time the
key will go into non-reversed video, and
simultaneously a
key will appear for inputting the reference value,
and
and
keys will appear for the upper and lower limit
values.
The settings of the reference value and the upper- and lower-limit values are
common to both the percentage mode and deviation percentage mode.
2.
Press the
key, and a screen will appear for setting the reference
value. This reference value is necessary for setting the limit values in terms
of percentages.
The setting can be made in the range
:
-200.00M to +200.00M
To return to the previous screen without making any change to the set
reference value, press the
key when the screen is in the state with
nothing being displayed (the state after pressing the
key).