38
────────────────────────────────────────────────────
4.3 Setting the Test Frequency
────────────────────────────────────────────────────
4.3.4 Input Using the Digit Screen
Shows the current
test frequency
The digit keys (If held down continuously, the corresponding digit increments
or decrements continuously.) (If the limit for the corresponding digit is reached,
these keys change to non-reversed video and become inactive.)
Changes the test
frequency to DC.
Monitor display portion
・
Test frequency
・
Test signal level
・
Voltage between test
samples
・
Limit value
・
Current flowing through
the test sample
Divides the test frequency by 10
Multiplies the test
frequency by 10
NOTE
1.
Access to the following Digit screen.
The current situation can be checked using the monitor display.
2.
Change the test frequency using the digit keys.
Frequency range which can be set: 0.000 Hz (DC) to 100.0 kHz
If you try to go out of this range, the relevant digit key goes into non-
reversed video and becomes inactive.
When measuring at 0 Hz (DC), and if the display parameter is set to
R
s or
R
p, the direct resistance (DCR) is displayed. The measured values for other
parameters are not displayed.
3.
When the setting has been established as above, press the
key to return
to the Initial screen.
Содержание 3522-50
Страница 1: ...Instruction Manual For 3522 50 LCR HiTESTER June 2011 Revised edition 7 3522C981 07 11 06H...
Страница 2: ......
Страница 8: ......
Страница 20: ...6 1 3 Names and Functions of Parts...
Страница 170: ...156 5 9 9442 PRINTER option...
Страница 197: ...183 7 5 Test Accuracy Conversion Table from C and L to Z Frequency Hz...
Страница 198: ...184 7 5 Test Accuracy...
Страница 201: ......
Страница 202: ......
Страница 203: ......
Страница 204: ......