78
────────────────────────────────────────────────────
4.12 Setting Averaging
────────────────────────────────────────────────────
4.12.2 Details of the Setting Process
Initial screen
Menu screen
The selected number is displayed, and averaging is set.
When the setting is completed, press this key
to return to the Initial screen.
Averaging Setting Screen
NOTE
1.
Press the
key on the Initial screen to display the Menu screen.
2.
Press the
key on the Menu screen to display the Averaging setting
screen.
3.
Press the key corresponding to the desired number of times for averaging.
The number of times for averaging which can be selected : 2, 4, 8, 16, 32, or
64.
The key which is pressed goes into non-reversed video, and this number is
set as the number of times for averaging.
The way in which averaging is performed varies, depending upon the trigger
setting.
With internal trigger:
A rolling average of the tested values over the set number of times for
averaging is always calculated backwards from the present.
When the sample to be tested is changed over, it takes a little time for a
certain stabilization time period until the results is reliable.
With external trigger:
An average of the test values is calculated over the set number of times for
averaging forwards from when the trigger is input.
Содержание 3522-50
Страница 1: ...Instruction Manual For 3522 50 LCR HiTESTER June 2011 Revised edition 7 3522C981 07 11 06H...
Страница 2: ......
Страница 8: ......
Страница 20: ...6 1 3 Names and Functions of Parts...
Страница 170: ...156 5 9 9442 PRINTER option...
Страница 197: ...183 7 5 Test Accuracy Conversion Table from C and L to Z Frequency Hz...
Страница 198: ...184 7 5 Test Accuracy...
Страница 201: ......
Страница 202: ......
Страница 203: ......
Страница 204: ......