58
────────────────────────────────────────────────────
4.7 Open Circuit Compensation
────────────────────────────────────────────────────
Open circuit
residual
component
Phase angle
NOTE
4.
When you press the
key, the SPOT open circuit compensation process
starts.
Interrupting the SPOT compensation process:
To interrupt the SPOT compensation process, press the
key on this
screen. The compensation process will stop and the display will revert to the
Initial screen. At this time, the compensation values obtained in the
previous episode of open circuit compensation will remain valid.
5.
If the compensation process has terminated satisfactorily, the following
Confirmation screen will be displayed.
The open circuit residual component for the test cables at the set frequency,
and its phase angle, are shown on the screen. (If an error message appears,
refer to Section 4.7.5.)
These values are required for use with the test range on HOLD. (For details,
refer to Section 4.9, "Open Circuit Compensation and Short Circuit
Compensation".)
After checking the open circuit residual component and its phase angle,
press the
key, and the display will return to the Initial screen.
Compensation can be performed for impedances of at least 1 k
Ω
.
Содержание 3522-50
Страница 1: ...Instruction Manual For 3522 50 LCR HiTESTER June 2011 Revised edition 7 3522C981 07 11 06H...
Страница 2: ......
Страница 8: ......
Страница 20: ...6 1 3 Names and Functions of Parts...
Страница 170: ...156 5 9 9442 PRINTER option...
Страница 197: ...183 7 5 Test Accuracy Conversion Table from C and L to Z Frequency Hz...
Страница 198: ...184 7 5 Test Accuracy...
Страница 201: ......
Страница 202: ......
Страница 203: ......
Страница 204: ......