3.3 Basic Measurement
────────────────────────────────────────────────────
Confirmation Screen
Initial Screen
7.
If the compensation process has terminated
satisfactorily, a Confirmation screen as shown to
the left is displayed.
The residual component at the frequency of 100
kHz with the test cables separated (open
circuited), and its phase angle, are shown on the
screen. These values are required for use with
the test range on HOLD.
(For details, refer to Section 4.9, "Open Circuit
Compensation and Short Circuit
Compensation".)
8.
After checking the open circuit residual
component and its phase angle, press the
key, and the display will return to the Initial
screen.
For details, refer to Section 4.7, "Open Circuit Compensation".