168
────────────────────────────────────────────────────
7.1 General Specification
────────────────────────────────────────────────────
14. Scaling function
Test value compensation is possible by setting the compensation
coefficients a and b to the following equation.
[test value after compensation] = a x [test value] + b
When the parameter is D or Q, calculate from compensation value
for
θ
.
15. Magnification display
function
The test values and comparator decision results can be displayed
in enlarged form
16. Continuous test
function
Continuous testing of the conditions which have been saved
17. Display digits
The values 3, 4, and 5 can be set as the number of displayed
digits. The actual number of displayed digits, however, depends
on the parameter used.
Parameter
Setting value
θ
, Q
Comparator
Δ
%
setting
D
Others
5
Up to the second
decimal place
Up to the fifth
decimal place
Up to 5 digits *1
4
Up to the first
decimal place
Up to the fourth
decimal place
Up to 4 digits
3
No decimal place
Up to the third
decimal place
Up to 3 digits
*1 When the measurement speed is set to FAST, the setting
is for 4 digits.
18. Display-setting
function
A setting can be made to turn on/off the indications of the LCD
display and voltage/current monitors.
19. System reset
All of the settings will revert to the factory settings.
20. Key lock function
(KEY LOCK)
This disables the operation of keys on the panel, by means of
either the special-purpose switch on the rear panel.
21. Panel save (SAVE)
and load (LOAD)
function
A maximum of 30 sets of complete measurement conditions can
be saved and recalled.
22. Beep sound setting
(BEEP)
There are settings corresponding to key input and to the
comparator results.
23. Interfaces
1) Touch panel settings
2) 9518-01 GP-IB INTERFACE or 9593-01 RS-232C
INTERFACE options can be selected (separate package)
3) EXT I/O connector (standard)
24. Temperature range
for use
0 to 40
(32
to 104
), up to 80% relative humidity, no
condensation.
25. Temperature range
for storage
-10 to 55
(14
to 131
), up to 80% relative humidity, no
condensation.
26. Operating
environment
Indoors, altitude up to 2000 m (6562 feet)
27. Power supply
Rated supply voltage 100/120/220/240 V AC (switchable)
(Voltage fluctuations of
10% from the
rated supply voltage are taken into
account.)
Rated supply frequency 50/60 Hz
Maximum rated power 40 VA (with options installed)
Содержание 3522-50
Страница 1: ...Instruction Manual For 3522 50 LCR HiTESTER June 2011 Revised edition 7 3522C981 07 11 06H...
Страница 2: ......
Страница 8: ......
Страница 20: ...6 1 3 Names and Functions of Parts...
Страница 170: ...156 5 9 9442 PRINTER option...
Страница 197: ...183 7 5 Test Accuracy Conversion Table from C and L to Z Frequency Hz...
Страница 198: ...184 7 5 Test Accuracy...
Страница 201: ......
Страница 202: ......
Страница 203: ......
Страница 204: ......