28
────────────────────────────────────────────────────
3.3 Basic Measurement
────────────────────────────────────────────────────
Confirmation Screen
Initial Screen
3.3.7 Starting Testing
7.
If the compensation process has terminated
satisfactorily, a Confirmation screen as shown
to the left is displayed.
The residual component at the frequency of
100 kHz with the test cables connected
together (short circuited), and its phase angle,
are shown on the screen.
These values are required for use with the test
range on HOLD. (For details, refer to Section
4.9, "Open Circuit Compensation and Short
Circuit Compensation".)
8.
After checking the short circuit residual
component and its phase angle, press the
key, and the display will return to the Initial
screen.
For details, refer to Section 4.8, "Short Circuit Compensation"
When the operations described above have been completed, the test
conditions have been for example set as follows:
Test conditions
Frequency 120 Hz
Constant voltage level 0.4 V
Open circuit compensation ALL compensation
Short circuit compensation ALL compensation
Parameters displayed
Capacitance Cs, Loss coefficient D
Connect the actual sample to be tested to the test cables, to start testing.