4.14 Setting and Activating the Comparator
────────────────────────────────────────────────────
N
umeric keypad
OFF key
U
nits are
percent (%)
NOTE
Percentage set value
= Reference value + |Reference value|
x
________________
100
Upper-limit
comparison value
3.
Select the units using one of the units keys:
:
step the units up
:
step the units down
p -- n --
μ
-- m -- (none) -- k -- M
<= Units down
Units up =>
Pressing the
key causes the upper limit value to be set and the screen
display to return to the Conditions setting screen.
4.
When the
key is pressed, a screen for setting the upper limit value
will appear. Use the numeric keypad to input the upper limit value as a
percentage relative to the reference value.
In order to set no upper limit value, press the
key.
The setting can be made in the range : -999.99% to +999.99%
The actual internal operation consists of calculating the upper-limit value of
comparison using the equation given below, and comparing it to the
measurement value to enable a decision to be made.
| |: Denotes an absolute value
When the setting has been established as above, press the
key to return
to the Conditions setting screen with the upper limit value set.
Содержание 3522-50
Страница 1: ...Instruction Manual For 3522 50 LCR HiTESTER June 2011 Revised edition 7 3522C981 07 11 06H...
Страница 2: ......
Страница 8: ......
Страница 20: ...6 1 3 Names and Functions of Parts...
Страница 170: ...156 5 9 9442 PRINTER option...
Страница 197: ...183 7 5 Test Accuracy Conversion Table from C and L to Z Frequency Hz...
Страница 198: ...184 7 5 Test Accuracy...
Страница 201: ......
Страница 202: ......
Страница 203: ......
Страница 204: ......