181
────────────────────────────────────────────────────
7.5 Test Accuracy
────────────────────────────────────────────────────
Zmin = 1.0144 k
Ω
×
(1 - 0.1/100
)≒
1.0134k
Ω
Zmax = 1.0144 k
Ω
×
(1 + 0.1/100
)≒
1.0154k
Ω
θ
min = 78.69 - 0.05 = 78.64
(
θ: |θ|
absolute value)
θ
max = 78.69 + 0.05 = 78.74
Csmin = 1/(Zmax
x
ω
x
sin
θ
max)
≒
159.81 nF ........-0.12%
Csmax = 1/(Zmin
x
ω
x
sin
θ
min)
≒
160.19 nF .......+0.12%
ω
= 2
x
π
x
f
f: frequency [Hz]
10 mV -
201 mV
501 mV -
1 V -
Coefficient
3
2
1.5
1
Test speed
FAST
NORMAL
SLOW
SLOW2
C
5
2
1.5
1
ON
OFF
Coefficient
4
1
Cable length
0 m
1 m
D
1
(100 kHz max) 1.5 + 0.015 x f [kHz]
(100.1 kHz min) 1.5 + 0.3 x f [MHz]
Temperature
Operating temperature = T
E
0.1 x basic accuracy x |T-23|
From the basic accuracy, find ranges that each of Z and
θ
can take.
From the ranges of Z and
θ,
determine ranges of Cs can take.
Hence the accuracy of Cs is
0.12
2. Test signal level (C)
3. Test speed coefficient (D)
4. DC bias coefficient (E)
5. Test cable length coefficient (F)
6. Temperature coefficient (G)
Содержание 3522-50
Страница 1: ...Instruction Manual For 3522 50 LCR HiTESTER June 2011 Revised edition 7 3522C981 07 11 06H...
Страница 2: ......
Страница 8: ......
Страница 20: ...6 1 3 Names and Functions of Parts...
Страница 170: ...156 5 9 9442 PRINTER option...
Страница 197: ...183 7 5 Test Accuracy Conversion Table from C and L to Z Frequency Hz...
Страница 198: ...184 7 5 Test Accuracy...
Страница 201: ......
Страница 202: ......
Страница 203: ......
Страница 204: ......