81
────────────────────────────────────────────────────
4.13 Setting the Testing Speed
────────────────────────────────────────────────────
4.13.2 Details of the Setting Process
Initial screen
Menu screen
The test speed currently set is displayed.
FAST key
NORM key
SLOW key
SLOW2 key
Test Speed Setting Screen
Testing speeds (typical)
FAST
5 ms
2 ms
Low accuracy testing at high speed
NORM
16 ms
2 ms
This speed used for normal testing
SLOW
88 ms
2 ms
High accuracy slow testing
SLOW2
828 ms
2 ms
High accuracy slow testing
1.
Press the
key on the Initial screen to display the Menu screen.
2.
Press the
key on the Menu screen to display the Test speed setting
screen.
3.
Select the desired new speed for testing from among the following options.
The key corresponding to the currently selected testing speed appears in non-
reversed video (black letters upon a white ground).
The testing speed varies according to the number of parameters being
displayed, and according to their type. The speeds given in the table relate
to the case of |
Z
| only being displayed. For details, refer to Section 7.3,
"Time Taken for Testing."
4.
When the setting has been established as above, press the
key to return
to the Initial screen.
Содержание 3522-50
Страница 1: ...Instruction Manual For 3522 50 LCR HiTESTER June 2011 Revised edition 7 3522C981 07 11 06H...
Страница 2: ......
Страница 8: ......
Страница 20: ...6 1 3 Names and Functions of Parts...
Страница 170: ...156 5 9 9442 PRINTER option...
Страница 197: ...183 7 5 Test Accuracy Conversion Table from C and L to Z Frequency Hz...
Страница 198: ...184 7 5 Test Accuracy...
Страница 201: ......
Страница 202: ......
Страница 203: ......
Страница 204: ......