Technical specifications|
iTernIS ADV
17.
12
Table 17-15
Ventilatory Parameters in Emergency Ventilation
Parameter
Initial Value
Range
Observations
Controlled pres-
sure (PCV)
10 cmH
2
O
2 – 30 cmH
2
O
The pressure is limited
by maximum pressure
alarm at 30 cmH
2
O.
This limit cannot be
changed.
Ti (inspiratory
time)
0.70 s
0.10 – 3.00 s
Inspiratory time may be
limited by the frequency
value.
Respiratory rate
20 rpm
1 – 100 rpm
The frequency may be
limited by the value of
the inspiratory time.
Insp. Detection
-2 cmH
2
O
-0.5 – 20
cmH
2
O
Detection by flow is not
enabled.
Concentration
of O
2
50
-
This parameter cannot
be changed.
Rise Time
Same value as
in PCV mode
Same range as
PCV mode
WARNING
• In the event that Emergency Ventilation is
activated, provide the patient with an alter
-
native ventilation mechanism, and find the
cause that generated this event. If it was
activated by the delay in accepting the first
initial calibration screen, restart the unit to
ventilate the patient.
• SAFETY VALVE: located immediately before the outlet to the patient. It opens when
the pressure within the patient circuit reaches a value of 120 ± 5 cmH
2
O. The gas
mixture passes into an internal collector and is discharged to the outside.
• INSPIRATORY RELIEF VALVE (ANTI-SUFFOCATION): this valve allows the pa
-
tient to inspire room air, when the unit is energized, or when the ventilator is in
inoperative status.
• OPERATING GAS LEAKAGE: The gas that can seep into the unit is collected in
a common manifold, and directed towards the outside.
• LACK OF COMPRESSED AIR PRESSURE: compressed air is used as a control
gas of the pneumatic circuit. When this is lacking, oxygen can replace it. Further-
more, when the unit detects lack of air, it automatically switches to an O
2
concen-
tration of 100%.
• LACK OF OXYGEN PRESSURE: in this case, the air takes the place of oxygen,
and the breaths are delivered with an O
2
concentration of 21%.
• AUTO-ZEROING: every 10 minutes, or when the operator enables it (by pressing
[Ctrl] + [Ctrl]
) all pressure sensors are zeroed. This restarts the readings, avoi-
ding offset errors of the measured pressures. During the first minutes of ventila
-
tion, the ventilator can implement resets with a frequency greater than the one
stated. At each reset, a message is displayed in the Icons and messages bar
indicating that this procedure is being performed.
Содержание iTernIS ADV
Страница 1: ...Operator s Manual Rev 1 0 07 2011 INNOVATIONS AS A SERVICE TO THE PATIENT ...
Страница 2: ......
Страница 3: ...User s Instruction Manual Rev 1 0 07 2011 ...
Страница 10: ...User Instruction Manual iTernIS ADV vi This Page Left Intentionally Blank ...
Страница 12: ...Introduction iTernIS ADV 1 2 This Page Left Intentionally Blank ...
Страница 20: ...Warnings Precautions Notes iTernIS ADV 2 2 This Page Left Intentionally Blank ...
Страница 24: ...Warnings Precautions Notes iTernIS ADV 2 6 This Page Left Intentionally Blank ...
Страница 26: ...Assembly and installation iTernIS ADV 3 2 This Page Left Intentionally Blank ...
Страница 38: ...General description Sections of the ventilator iTernIS ADV 4 2 This Page Left Intentionally Blank ...
Страница 50: ...Start up and initial calibration iTernIS ADV 5 2 This Page Left Intentionally Blank ...
Страница 60: ...Start up and initial calibration iTernIS ADV 5 12 This Page Left Intentionally Blank ...
Страница 62: ...Basic operations iTernIS ADV 6 2 This Page Left Intentionally Blank ...
Страница 74: ...Basic operations iTernIS ADV 6 14 This Page Left Intentionally Blank ...
Страница 76: ...Menu iTernIS ADV 7 2 This Page Left Intentionally Blank ...
Страница 92: ...Menu iTernIS ADV 7 18 This Page Left Intentionally Blank ...
Страница 94: ...Graphics management iTernIS ADV 8 2 This Page Left Intentionally Blank ...
Страница 100: ...Graphics management iTernIS ADV 8 8 This Page Left Intentionally Blank ...
Страница 102: ...ADL PED Operative modes iTernIS ADV 9 2 This Page Left Intentionally Blank ...
Страница 120: ...ADL PED Operative modes iTernIS ADV 9 20 This Page Left Intentionally Blank ...
Страница 122: ...NEO INF Operative modes iTernIS ADV 10 2 This Page Left Intentionally Blank ...
Страница 136: ...NEO INF Operative modes iTernIS ADV 10 16 This Page Left Intentionally Blank ...
Страница 138: ...Alarms iTernIS ADV 11 2 This Page Left Intentionally Blank ...
Страница 148: ...Alarms iTernIS ADV 11 12 This Page Left Intentionally Blank ...
Страница 150: ...Respiratory mechanics iTernIS ADV 12 2 This Page Left Intentionally Blank ...
Страница 162: ...Respiratory mechanics iTernIS ADV 12 14 Fig 12 8 VD VT physiologic test screen ...
Страница 164: ...Trends iTernIS ADV 13 2 This Page Left Intentionally Blank ...
Страница 168: ...Trends iTernIS ADV 13 6 This Page Left Intentionally Blank ...
Страница 170: ...Capnography iTernIS ADV 14 2 This Page Left Intentionally Blank ...
Страница 176: ...Capnography iTernIS ADV 14 8 This Page Left Intentionally Blank ...
Страница 178: ...Operational verification iTernIS ADV 15 2 This Page Left Intentionally Blank ...
Страница 182: ...Operational verification iTernIS ADV 15 6 This Page Left Intentionally Blank ...
Страница 184: ...Cleaning and maintenance iTernIS ADV 16 2 This Page Left Intentionally Blank ...
Страница 192: ...Technical specifications iTernIS ADV 17 2 This Page Left Intentionally Blank ...
Страница 205: ...Appendices iTernIS ADV A i Appendices Chapter Contents A I Troubleshootings ...
Страница 206: ...Appendices iTernIS ADV A ii This Page Left Intentionally Blank ...
Страница 212: ...Warranty iTernIS ADV G ii This Page Left Intentionally Blank ...
Страница 215: ...Alphabetical index iTernIS ADV I iii W Watchdog 17 9 WOBi 7 3 Y Y connector or part 3 8 ...
Страница 216: ...Alphabetical index iTernIS ADV I iv This Page Left Intentionally Blank ...
Страница 217: ......
Страница 218: ......