NEO-INF Operative modes |
iTernIS ADV
10.
12
• Manual inspiration.
• Nebulization.
• Volume compensation.
• Leak compensation.
• Distal/proximal pneumotachograph.
This mode has spontaneous ventilation and positive continuous airway pressure
for noninvasive application to neonatal-infant patients (using a nasal prong).
There is continuous (adjustable) flow during CPAP breaths. This mode does not
have ventilation support cycles. With each inspiratory effort, the patient receives
part of the gas mixture circulating as continuous flow, in a magnitude that is propor
-
tional to the intensity required. The ventilator monitors the presence of these res-
piratory efforts and, if it does not detect them, it assumes an apnea condition to be
present and enables Backup Ventilation, triggering the appropriate alarm signals.
In this mode, leakage compensation is enabled by default.
It must be understood that the selection of high continuous flows can raise the
pressure baseline due to the more or less constant resistance presented by the
entire respiratory system. Due to this resistence, a flow increase results in an in
-
crease in the difference in ingoing/outgoing pressure. Therefore, the ability of the
professional in charge to determine the PEEP and continuous flow levels is very
important.
WARNING
• In Continuous Flow CPAP ventilation, a com
-
petent professional must always be present
to control the patient and the ventilator. Be
-
cause of the high resistance of the nasal
prongs for NEO-INF patients and the large
leaks that usually occur in this ventilation
mode, nasal prong disconnection warning
conditions may not be detected.
NOTE
• The tidal volume and minute volume alarms are
disabled in this mode.
10.9 | Continuous Flow
CPAP
Содержание iTernIS ADV
Страница 1: ...Operator s Manual Rev 1 0 07 2011 INNOVATIONS AS A SERVICE TO THE PATIENT ...
Страница 2: ......
Страница 3: ...User s Instruction Manual Rev 1 0 07 2011 ...
Страница 10: ...User Instruction Manual iTernIS ADV vi This Page Left Intentionally Blank ...
Страница 12: ...Introduction iTernIS ADV 1 2 This Page Left Intentionally Blank ...
Страница 20: ...Warnings Precautions Notes iTernIS ADV 2 2 This Page Left Intentionally Blank ...
Страница 24: ...Warnings Precautions Notes iTernIS ADV 2 6 This Page Left Intentionally Blank ...
Страница 26: ...Assembly and installation iTernIS ADV 3 2 This Page Left Intentionally Blank ...
Страница 38: ...General description Sections of the ventilator iTernIS ADV 4 2 This Page Left Intentionally Blank ...
Страница 50: ...Start up and initial calibration iTernIS ADV 5 2 This Page Left Intentionally Blank ...
Страница 60: ...Start up and initial calibration iTernIS ADV 5 12 This Page Left Intentionally Blank ...
Страница 62: ...Basic operations iTernIS ADV 6 2 This Page Left Intentionally Blank ...
Страница 74: ...Basic operations iTernIS ADV 6 14 This Page Left Intentionally Blank ...
Страница 76: ...Menu iTernIS ADV 7 2 This Page Left Intentionally Blank ...
Страница 92: ...Menu iTernIS ADV 7 18 This Page Left Intentionally Blank ...
Страница 94: ...Graphics management iTernIS ADV 8 2 This Page Left Intentionally Blank ...
Страница 100: ...Graphics management iTernIS ADV 8 8 This Page Left Intentionally Blank ...
Страница 102: ...ADL PED Operative modes iTernIS ADV 9 2 This Page Left Intentionally Blank ...
Страница 120: ...ADL PED Operative modes iTernIS ADV 9 20 This Page Left Intentionally Blank ...
Страница 122: ...NEO INF Operative modes iTernIS ADV 10 2 This Page Left Intentionally Blank ...
Страница 136: ...NEO INF Operative modes iTernIS ADV 10 16 This Page Left Intentionally Blank ...
Страница 138: ...Alarms iTernIS ADV 11 2 This Page Left Intentionally Blank ...
Страница 148: ...Alarms iTernIS ADV 11 12 This Page Left Intentionally Blank ...
Страница 150: ...Respiratory mechanics iTernIS ADV 12 2 This Page Left Intentionally Blank ...
Страница 162: ...Respiratory mechanics iTernIS ADV 12 14 Fig 12 8 VD VT physiologic test screen ...
Страница 164: ...Trends iTernIS ADV 13 2 This Page Left Intentionally Blank ...
Страница 168: ...Trends iTernIS ADV 13 6 This Page Left Intentionally Blank ...
Страница 170: ...Capnography iTernIS ADV 14 2 This Page Left Intentionally Blank ...
Страница 176: ...Capnography iTernIS ADV 14 8 This Page Left Intentionally Blank ...
Страница 178: ...Operational verification iTernIS ADV 15 2 This Page Left Intentionally Blank ...
Страница 182: ...Operational verification iTernIS ADV 15 6 This Page Left Intentionally Blank ...
Страница 184: ...Cleaning and maintenance iTernIS ADV 16 2 This Page Left Intentionally Blank ...
Страница 192: ...Technical specifications iTernIS ADV 17 2 This Page Left Intentionally Blank ...
Страница 205: ...Appendices iTernIS ADV A i Appendices Chapter Contents A I Troubleshootings ...
Страница 206: ...Appendices iTernIS ADV A ii This Page Left Intentionally Blank ...
Страница 212: ...Warranty iTernIS ADV G ii This Page Left Intentionally Blank ...
Страница 215: ...Alphabetical index iTernIS ADV I iii W Watchdog 17 9 WOBi 7 3 Y Y connector or part 3 8 ...
Страница 216: ...Alphabetical index iTernIS ADV I iv This Page Left Intentionally Blank ...
Страница 217: ......
Страница 218: ......