Technical specifications |
iTernIS ADV
17.
7
Apnea (s)
5 - 60
5
15
Low PEEP
(cmH
2
O)
0 - 6
2
ADL/PED/NEO-
INF: 2
Maximum respi-
ratory rate
(rpm)
3 - 160
1
ADL/PED: 30
NEO-INF: 40
ETCO
2
(mmHg)
Max: >Min to 150
Min 1 to <Max
1
Max: 50
Min: 30
Continuous high
pressure
Alarm condition: 5 cmH
2
O above the configured PEEP
level for over 15 s. Not configurable.
External power
failure
Not configurable. Whatever the cause of the event may be,
the ventilator automatically switches to battery operation.
Dead battery
Null or very low charge level. Replace by an alternative
respiratory method. Not configurable.
Inop. Vent.
When the operation of the ventilator is suspended, replace
by an alternative ventilatory method. Not configurable.
Blower failure
Not configurable. Alarm signals are triggered when exces
-
sive power consumption by the blower is detected (caused
by increased resistance at the rotation of the blades).
(1) From 1 to 30 mL it increases per 0.001 L; from 30 to 100 mL it increases per
0.005 L, and >100 mL it increases per 0.010 L.
(2) In modes of control or target volume, the VTmax is 1.5 IBW-based VT, and
the initial VTmin is 0.5 IBW-based VT .
(3) The values declared for ADL and PED correspond to the NIV mode. For the
rest of the modes the Leakage out of range is triggered starting at 15 L/min. In
NEO-INF the leakage within range is the same for all modes.
Table 17-10
Data collection for control and monitoring
Parameter
Method
Involved Variables
Accuracy
Accuracy
Relative pressure
transducers
Control, support and
monitored pressures
±2 cmH
2
O
Controlled
flow
Mesh pneumotachogra-
ph related to differential
pressure transducers
Continuous flow, flow
for controlled volume
generation
±10%
Monitored
flow
Variable opening
pneumotachograph
related to the pressure
differential transducer
(optional fixed opening
pneumotachograph for
the NEO -INF category)
Monitored and plotted
flows, and expired
volumes
±10%
Volume
It is derived from the
corresponding flow
measurements
Controlled, monitored
and plotted volumes
±10%
Time
Quartz crystal.
Inspiratory and expira-
tory times, respiratory
rate, and I:E ratio.
Internal clock.
±0,06 s
± 1 rpm
Concentra-
tion of O
2
Galvanic cell
Concentration of O
2
.
±3% Vol.
WARNING
• The replacement of cables (main power su
-
pply or internal cabling) of the equipment by
anyone other than those approved by HEYER
may result in decreased immunity, or increa
-
sed emissions reported for iTernIS ADV.
17.10 | Data Collection
for Control and
Monitoring
17.11 | Electromagnetic
Compatibility
Содержание iTernIS ADV
Страница 1: ...Operator s Manual Rev 1 0 07 2011 INNOVATIONS AS A SERVICE TO THE PATIENT ...
Страница 2: ......
Страница 3: ...User s Instruction Manual Rev 1 0 07 2011 ...
Страница 10: ...User Instruction Manual iTernIS ADV vi This Page Left Intentionally Blank ...
Страница 12: ...Introduction iTernIS ADV 1 2 This Page Left Intentionally Blank ...
Страница 20: ...Warnings Precautions Notes iTernIS ADV 2 2 This Page Left Intentionally Blank ...
Страница 24: ...Warnings Precautions Notes iTernIS ADV 2 6 This Page Left Intentionally Blank ...
Страница 26: ...Assembly and installation iTernIS ADV 3 2 This Page Left Intentionally Blank ...
Страница 38: ...General description Sections of the ventilator iTernIS ADV 4 2 This Page Left Intentionally Blank ...
Страница 50: ...Start up and initial calibration iTernIS ADV 5 2 This Page Left Intentionally Blank ...
Страница 60: ...Start up and initial calibration iTernIS ADV 5 12 This Page Left Intentionally Blank ...
Страница 62: ...Basic operations iTernIS ADV 6 2 This Page Left Intentionally Blank ...
Страница 74: ...Basic operations iTernIS ADV 6 14 This Page Left Intentionally Blank ...
Страница 76: ...Menu iTernIS ADV 7 2 This Page Left Intentionally Blank ...
Страница 92: ...Menu iTernIS ADV 7 18 This Page Left Intentionally Blank ...
Страница 94: ...Graphics management iTernIS ADV 8 2 This Page Left Intentionally Blank ...
Страница 100: ...Graphics management iTernIS ADV 8 8 This Page Left Intentionally Blank ...
Страница 102: ...ADL PED Operative modes iTernIS ADV 9 2 This Page Left Intentionally Blank ...
Страница 120: ...ADL PED Operative modes iTernIS ADV 9 20 This Page Left Intentionally Blank ...
Страница 122: ...NEO INF Operative modes iTernIS ADV 10 2 This Page Left Intentionally Blank ...
Страница 136: ...NEO INF Operative modes iTernIS ADV 10 16 This Page Left Intentionally Blank ...
Страница 138: ...Alarms iTernIS ADV 11 2 This Page Left Intentionally Blank ...
Страница 148: ...Alarms iTernIS ADV 11 12 This Page Left Intentionally Blank ...
Страница 150: ...Respiratory mechanics iTernIS ADV 12 2 This Page Left Intentionally Blank ...
Страница 162: ...Respiratory mechanics iTernIS ADV 12 14 Fig 12 8 VD VT physiologic test screen ...
Страница 164: ...Trends iTernIS ADV 13 2 This Page Left Intentionally Blank ...
Страница 168: ...Trends iTernIS ADV 13 6 This Page Left Intentionally Blank ...
Страница 170: ...Capnography iTernIS ADV 14 2 This Page Left Intentionally Blank ...
Страница 176: ...Capnography iTernIS ADV 14 8 This Page Left Intentionally Blank ...
Страница 178: ...Operational verification iTernIS ADV 15 2 This Page Left Intentionally Blank ...
Страница 182: ...Operational verification iTernIS ADV 15 6 This Page Left Intentionally Blank ...
Страница 184: ...Cleaning and maintenance iTernIS ADV 16 2 This Page Left Intentionally Blank ...
Страница 192: ...Technical specifications iTernIS ADV 17 2 This Page Left Intentionally Blank ...
Страница 205: ...Appendices iTernIS ADV A i Appendices Chapter Contents A I Troubleshootings ...
Страница 206: ...Appendices iTernIS ADV A ii This Page Left Intentionally Blank ...
Страница 212: ...Warranty iTernIS ADV G ii This Page Left Intentionally Blank ...
Страница 215: ...Alphabetical index iTernIS ADV I iii W Watchdog 17 9 WOBi 7 3 Y Y connector or part 3 8 ...
Страница 216: ...Alphabetical index iTernIS ADV I iv This Page Left Intentionally Blank ...
Страница 217: ......
Страница 218: ......