Instrument Configuration
89
•
•
•
•
•
•
On the HI 6500 series instrument display
, select the communication menu and use the
UP or DOWN button to select the Printer submenu, then select the Baud Rate menu item.
The current Baud Rate value will be displayed, press the ENTER or the RIGHT button if
the value needs to be modified.
Parity Bit Parameter
The Parity Bit parameter has three options, EVEN, ODD, or OFF.
On the Web page
, select the communication menu, and then select the Printer submenu,
and left click inside the Parity field and select between EVEN, ODD, or OFF using the drop
down menu..
On the HI 6500 series instrument display
, select the communication menu and use
the UP or DOWN button to select the Printer submenu, then select the Parity menu
item. The current Parity value will be displayed, press the ENTER or the RIGHT
button if the Parity needs to be modified.
USB Memory Stick Parameters
The HI 6500 provides an interface to an external USB memory stick that allows parameters
to be saved, restored or transfered to another instrument.
Restore Command
This command replaces the existing HI 6500 series parameters by the parameters stored on
a USB memory stick.
You can create a "Restore.txt" file on the USB drive that when present in the HI 6500
directory, upon powerup, the unit will perform a restore function using this file and
overwriting the affected parameters in the unit.
You can also create a safety "restore.txt" file by using a Params.txt file created when
doing a "SAVE" function to the USB drive and copying/renaming it to Restore.txt.
This will enable you to modify this Restore file by making any required changes or
deleting parameters or changing parameter values in the file. Any changed values will
overwrite parameters in the unit.
On the Web page
, go the communication menu and left click on the Restore button.
Содержание HI 6500 Series
Страница 10: ...viii Contents ...
Страница 20: ...10 Chapter 2 ...
Страница 72: ...62 Chapter 4 ...
Страница 96: ...86 Chapter 5 Profibus DP Commands and Parameters ...
Страница 97: ...Instrument Configuration 87 Serial Option Termination ...
Страница 126: ...116 Chapter 5 ...
Страница 156: ...146 Chapter 8 ...
Страница 163: ...Troubleshooting 153 General Troubleshooting Flow Chart Index ...
Страница 164: ...154 Chapter 9 A Guideline Instability Electrical and Mechanical review ...
Страница 165: ...Troubleshooting 155 A1 Checking for Unstable Components in a Weighing System ...
Страница 166: ...156 Chapter 9 B Guidelines for Electrical Mechanical or Configuration Issues ...
Страница 167: ...Troubleshooting 157 B1 Guidelines to Verify Electrical Installation ...
Страница 168: ...158 Chapter 9 B2 Guidelines to Verify Mechanical Installation ...
Страница 169: ...Troubleshooting 159 B3 Guidelines to Verify Configuration Filter Settings to Improve Stability ...
Страница 170: ...160 Chapter 9 C Integrated Technician and Stability Test Overview ...
Страница 171: ...Troubleshooting 161 E Testing for Non Return to Zero System with IT Summing Card ...
Страница 173: ...Troubleshooting 163 G Calibration Errors When Performing the Calibration ...
Страница 174: ...164 Chapter 9 H Mechanical Installation ...
Страница 175: ...Troubleshooting 165 J Electrical Inspection ...
Страница 176: ...166 Chapter 9 K Installation Check Points ...
Страница 177: ...Troubleshooting 167 M Weight Processor s Front Display is Blank or Locked ...
Страница 178: ...168 Chapter 9 N Analog Out HI 6510 Only ...
Страница 194: ...184 Appendix A Figure D1 Assembling the Swivel Mount Enclosure on the instrument pedestal ...
Страница 195: ...Installation of Swivel Mount Enclosures 185 Figure D2 Assembling the Swivel Mount Enclosure on a Table ...
Страница 196: ...186 Appendix A Dimensions ...
Страница 200: ...4 HI 6500 Series USER GUIDE Index ...
Страница 201: ......