Troubleshooting
173
•
•
•
•
•
•
Running the Stability Test from the Web Interface
A Stability Test column on the IT test results display (see picture above) shows PASS or
FAIL for each load sensor. We obtained the data by running the IT test, as described above.
Stability test is in the Diagnostics menu. Select the Stability Test hyperlink and wait a few
seconds.
The results will not be specific to each load sensor, as in the IT test. They will display an
average reading and PASS/FAIL for the system as a whole.If equipped with an IT summing
junction box there will be a selection to run the Weight and voltage test. That test will
display the individual load cell and loop back test results.
Running the Stability Test from the front panel
Step 1. From the Diagnostic menu, select Stability Test with use the down arrow and press
Enter. The HI 6500 runs the test and shows the results.
Step 2. To see the AD/Raw Count and the A/D Filtered select them from the menu.
Step 3. Press Exit to return to the Test Menu.
WARNING - D
O
NOT
PERFORM
THE
S
TABILITY
T
EST
DURING
PRODUCTION
. T
HE
TEST
ACTIVITIES
CAN
CAUSE
INCORRECT
READINGS
.
AVERTISSEMENT – Ne pas effectuer le test de stabilité lors de la production.
Ces tests peuvent résulter à des lectures incorrectes.
Содержание HI 6500 Series
Страница 10: ...viii Contents ...
Страница 20: ...10 Chapter 2 ...
Страница 72: ...62 Chapter 4 ...
Страница 96: ...86 Chapter 5 Profibus DP Commands and Parameters ...
Страница 97: ...Instrument Configuration 87 Serial Option Termination ...
Страница 126: ...116 Chapter 5 ...
Страница 156: ...146 Chapter 8 ...
Страница 163: ...Troubleshooting 153 General Troubleshooting Flow Chart Index ...
Страница 164: ...154 Chapter 9 A Guideline Instability Electrical and Mechanical review ...
Страница 165: ...Troubleshooting 155 A1 Checking for Unstable Components in a Weighing System ...
Страница 166: ...156 Chapter 9 B Guidelines for Electrical Mechanical or Configuration Issues ...
Страница 167: ...Troubleshooting 157 B1 Guidelines to Verify Electrical Installation ...
Страница 168: ...158 Chapter 9 B2 Guidelines to Verify Mechanical Installation ...
Страница 169: ...Troubleshooting 159 B3 Guidelines to Verify Configuration Filter Settings to Improve Stability ...
Страница 170: ...160 Chapter 9 C Integrated Technician and Stability Test Overview ...
Страница 171: ...Troubleshooting 161 E Testing for Non Return to Zero System with IT Summing Card ...
Страница 173: ...Troubleshooting 163 G Calibration Errors When Performing the Calibration ...
Страница 174: ...164 Chapter 9 H Mechanical Installation ...
Страница 175: ...Troubleshooting 165 J Electrical Inspection ...
Страница 176: ...166 Chapter 9 K Installation Check Points ...
Страница 177: ...Troubleshooting 167 M Weight Processor s Front Display is Blank or Locked ...
Страница 178: ...168 Chapter 9 N Analog Out HI 6510 Only ...
Страница 194: ...184 Appendix A Figure D1 Assembling the Swivel Mount Enclosure on the instrument pedestal ...
Страница 195: ...Installation of Swivel Mount Enclosures 185 Figure D2 Assembling the Swivel Mount Enclosure on a Table ...
Страница 196: ...186 Appendix A Dimensions ...
Страница 200: ...4 HI 6500 Series USER GUIDE Index ...
Страница 201: ......