54
Chapter
4
•
•
•
•
•
•
Step 3. Cycle power, or perform the two previous steps before powering up your network.
Address selection can only occur after cycling the power. Make sure that the
software you use will detect the values as you have set them.
Step 4. Complete any additional configuration that is required by your PLC for
initialization. Our initialization example is for a Siemens PLC. Your PLC
initialization requirements may differ.
Step 5. Install the *.GSD file for the instrument you connected to the Profibus
Network.
Configuring Profibus from the Front Panel
Step 1. Press the Configuration key
Step 2. Down arrow to Communications; press enter.
Step 3. Select Profibus-DP; press enter.
Step 4. Select Termination; press enter
Step 5. Up or down arrow to make termination selection. Press enter to select.
Step 6. Down arrow to select Serial Option;
Step 7. Press enter to toggle between Profibus-DP and Modbus. Select Profibus-DP and
press enter.
Step 8. Down arrow to Node.
Step 9. Press enter to make changes to the Node. Use the arrow keys to enter node number
and press enter to accept.
Step 10. Press exit key until display returns to the Summary Display.
Initialization Procedures on the PLC
NOTE
The examples come from the Siemens Step 7™, Simatic Manager Software. Your software
will vary from these procedures. Step 7™ is a trademark of the Siemens Corporation.
Step 1. Step 1. In the Siemens Step 7™, Simatic Manager, open the Hardware Catalog.
CAUTION: The address should never be changed during operation. if the address is
modified while on line, an internal error could be generated and the instrument
disconnected from the network.
ATTENTION L’adresse ne devrait jamaisêtre changée durant le
fonctionnement. Si l’adresse est modifiée pendant la miseen ligne, une erreur
interne pourrait être créée et le module sera déconnecté du réseau.
Содержание HI 6500 Series
Страница 10: ...viii Contents ...
Страница 20: ...10 Chapter 2 ...
Страница 72: ...62 Chapter 4 ...
Страница 96: ...86 Chapter 5 Profibus DP Commands and Parameters ...
Страница 97: ...Instrument Configuration 87 Serial Option Termination ...
Страница 126: ...116 Chapter 5 ...
Страница 156: ...146 Chapter 8 ...
Страница 163: ...Troubleshooting 153 General Troubleshooting Flow Chart Index ...
Страница 164: ...154 Chapter 9 A Guideline Instability Electrical and Mechanical review ...
Страница 165: ...Troubleshooting 155 A1 Checking for Unstable Components in a Weighing System ...
Страница 166: ...156 Chapter 9 B Guidelines for Electrical Mechanical or Configuration Issues ...
Страница 167: ...Troubleshooting 157 B1 Guidelines to Verify Electrical Installation ...
Страница 168: ...158 Chapter 9 B2 Guidelines to Verify Mechanical Installation ...
Страница 169: ...Troubleshooting 159 B3 Guidelines to Verify Configuration Filter Settings to Improve Stability ...
Страница 170: ...160 Chapter 9 C Integrated Technician and Stability Test Overview ...
Страница 171: ...Troubleshooting 161 E Testing for Non Return to Zero System with IT Summing Card ...
Страница 173: ...Troubleshooting 163 G Calibration Errors When Performing the Calibration ...
Страница 174: ...164 Chapter 9 H Mechanical Installation ...
Страница 175: ...Troubleshooting 165 J Electrical Inspection ...
Страница 176: ...166 Chapter 9 K Installation Check Points ...
Страница 177: ...Troubleshooting 167 M Weight Processor s Front Display is Blank or Locked ...
Страница 178: ...168 Chapter 9 N Analog Out HI 6510 Only ...
Страница 194: ...184 Appendix A Figure D1 Assembling the Swivel Mount Enclosure on the instrument pedestal ...
Страница 195: ...Installation of Swivel Mount Enclosures 185 Figure D2 Assembling the Swivel Mount Enclosure on a Table ...
Страница 196: ...186 Appendix A Dimensions ...
Страница 200: ...4 HI 6500 Series USER GUIDE Index ...
Страница 201: ......