174
Chapter
9
•
•
•
•
•
•
Weight and Voltage Tests
The Weight and Voltage tests are used to diagnose a weighing system and, if certain types
of problems are indicated, determine their source. It provides the total scale input to the
instrument such as mV, mV/V and Weight in the units selected (i.e. lbs, kg, oz, g).
Weight and Voltage Test from the Web interface
Since the IT Web page shows all the weight and voltage values at once, it is the preferred
method for troubleshooting. Click Integrated Technician to open the Diagnostics -
Integrated Technician page.
To view individual load sensor data, click Do IT Test.
These readings are snapshots and allow you to determine if the problem is in the instrument
(internal) or in a load sensor(s) (external).
The specification range for the Weight Processor is 0-25 mV, so a reading between 0-25 mV
is within the normal range. A reading outside this range would normally indicate that the
problem is external to the system (most likely improper wiring).
Stability: Gross = 0.9lb
RAW
Waversaver
Mean 32918
32993
Variation
0.17
0.00
Results
OK
OK
mV/V 0.0184
RTZ Pass
Содержание HI 6500 Series
Страница 10: ...viii Contents ...
Страница 20: ...10 Chapter 2 ...
Страница 72: ...62 Chapter 4 ...
Страница 96: ...86 Chapter 5 Profibus DP Commands and Parameters ...
Страница 97: ...Instrument Configuration 87 Serial Option Termination ...
Страница 126: ...116 Chapter 5 ...
Страница 156: ...146 Chapter 8 ...
Страница 163: ...Troubleshooting 153 General Troubleshooting Flow Chart Index ...
Страница 164: ...154 Chapter 9 A Guideline Instability Electrical and Mechanical review ...
Страница 165: ...Troubleshooting 155 A1 Checking for Unstable Components in a Weighing System ...
Страница 166: ...156 Chapter 9 B Guidelines for Electrical Mechanical or Configuration Issues ...
Страница 167: ...Troubleshooting 157 B1 Guidelines to Verify Electrical Installation ...
Страница 168: ...158 Chapter 9 B2 Guidelines to Verify Mechanical Installation ...
Страница 169: ...Troubleshooting 159 B3 Guidelines to Verify Configuration Filter Settings to Improve Stability ...
Страница 170: ...160 Chapter 9 C Integrated Technician and Stability Test Overview ...
Страница 171: ...Troubleshooting 161 E Testing for Non Return to Zero System with IT Summing Card ...
Страница 173: ...Troubleshooting 163 G Calibration Errors When Performing the Calibration ...
Страница 174: ...164 Chapter 9 H Mechanical Installation ...
Страница 175: ...Troubleshooting 165 J Electrical Inspection ...
Страница 176: ...166 Chapter 9 K Installation Check Points ...
Страница 177: ...Troubleshooting 167 M Weight Processor s Front Display is Blank or Locked ...
Страница 178: ...168 Chapter 9 N Analog Out HI 6510 Only ...
Страница 194: ...184 Appendix A Figure D1 Assembling the Swivel Mount Enclosure on the instrument pedestal ...
Страница 195: ...Installation of Swivel Mount Enclosures 185 Figure D2 Assembling the Swivel Mount Enclosure on a Table ...
Страница 196: ...186 Appendix A Dimensions ...
Страница 200: ...4 HI 6500 Series USER GUIDE Index ...
Страница 201: ......