Calibration
121
•
•
•
•
•
•
C2 and eCAL Calibration from the Web Page
From the Home Page, click Configuration to display the Configuration menu; then
click Calibration to open the Calibration page.
Step 1. Check to make sure the number of load cells displayed on the webpage
corresponds correctly to the inputs on the instrument.
Step 2. In the “Ref Weight” text field, enter the reference weight for your application.
Step 3. An empty scale will use a 0.00 reference weight setting.
Step 4. (Optional) If you want to enter the Gravity Correction Factor, see the table
above, then enter the Correction Factor number you selected from the table in
the Gravity Correction field
Step 5. Click the “Do C2 Calibration” button
Step 6. Wait a few seconds and the results will appear. If the Calibration was successful
a Cal completed OK appears and gross weight will read your reference value.
However, if no load points are connected to the HI 6500 series the following
message appears. Check the cable and connectors of the load point(s) and re-
calibrate.
Step 7. This message appears if no load points are connected to the HI 6500. Check the
cable and connectors of the load point(s) and re-calibrate.
Cal failed: A to D converter error
Содержание HI 6500 Series
Страница 10: ...viii Contents ...
Страница 20: ...10 Chapter 2 ...
Страница 72: ...62 Chapter 4 ...
Страница 96: ...86 Chapter 5 Profibus DP Commands and Parameters ...
Страница 97: ...Instrument Configuration 87 Serial Option Termination ...
Страница 126: ...116 Chapter 5 ...
Страница 156: ...146 Chapter 8 ...
Страница 163: ...Troubleshooting 153 General Troubleshooting Flow Chart Index ...
Страница 164: ...154 Chapter 9 A Guideline Instability Electrical and Mechanical review ...
Страница 165: ...Troubleshooting 155 A1 Checking for Unstable Components in a Weighing System ...
Страница 166: ...156 Chapter 9 B Guidelines for Electrical Mechanical or Configuration Issues ...
Страница 167: ...Troubleshooting 157 B1 Guidelines to Verify Electrical Installation ...
Страница 168: ...158 Chapter 9 B2 Guidelines to Verify Mechanical Installation ...
Страница 169: ...Troubleshooting 159 B3 Guidelines to Verify Configuration Filter Settings to Improve Stability ...
Страница 170: ...160 Chapter 9 C Integrated Technician and Stability Test Overview ...
Страница 171: ...Troubleshooting 161 E Testing for Non Return to Zero System with IT Summing Card ...
Страница 173: ...Troubleshooting 163 G Calibration Errors When Performing the Calibration ...
Страница 174: ...164 Chapter 9 H Mechanical Installation ...
Страница 175: ...Troubleshooting 165 J Electrical Inspection ...
Страница 176: ...166 Chapter 9 K Installation Check Points ...
Страница 177: ...Troubleshooting 167 M Weight Processor s Front Display is Blank or Locked ...
Страница 178: ...168 Chapter 9 N Analog Out HI 6510 Only ...
Страница 194: ...184 Appendix A Figure D1 Assembling the Swivel Mount Enclosure on the instrument pedestal ...
Страница 195: ...Installation of Swivel Mount Enclosures 185 Figure D2 Assembling the Swivel Mount Enclosure on a Table ...
Страница 196: ...186 Appendix A Dimensions ...
Страница 200: ...4 HI 6500 Series USER GUIDE Index ...
Страница 201: ......