Network Configuration
37
•
•
•
•
•
•
A 32-bit value,
Instrument Status
, provides the current state of all the major
functions within the
instrument. The top 8 bits are a cyclic “measurement
update count”, which will increment by a count of one every time a new
measurement value is taken, following a 0 to 255 then repeat cycle. If this
value remains the same in two consecutive reads from the instrument then
the communication or the measurement function has failed and the
appropriate action needs to be taken. The bottom 24-bits reflect the status
of all the major functions and should be used in conjunction with the
“measurement update count” to determine the health of the instrument.
The
Net
and
Gross Weight
values are always provided.
The final five read only command values
Parameter RD1 Value
,
Parameter RD2 Value
,
Parameter RD3 Value
,
Parameter RD4 Value
, and
Parameter RD5 Value
are the read
only values for the different user selectable parameter ID values set in the output table. If a
0x0000 is placed in the
Parameter RDx ID value
, or the requested parameter does not
exist a 0x0000 value will be returned.
Hardy Command Numbers
Here is a list of Hardy command numbers:
Command number
Command
0
Read Parameter
1
Zero Cmd
2
Tare Cmd
4
Write Non-Volatile Cmd
5
Print Cmd
6
Weigh Sample Cmd
0x64 (100 dec)
Cal Low Cmd
0x65 (101 dec)
Cal High Cmd
0x66 (102 dec)
C2 Cal Cmd
0x1000 (4096 dec)
Write Integer Cmd
0x1001 (4097 dec)
Write Float Cmd
Содержание HI 6500 Series
Страница 10: ...viii Contents ...
Страница 20: ...10 Chapter 2 ...
Страница 72: ...62 Chapter 4 ...
Страница 96: ...86 Chapter 5 Profibus DP Commands and Parameters ...
Страница 97: ...Instrument Configuration 87 Serial Option Termination ...
Страница 126: ...116 Chapter 5 ...
Страница 156: ...146 Chapter 8 ...
Страница 163: ...Troubleshooting 153 General Troubleshooting Flow Chart Index ...
Страница 164: ...154 Chapter 9 A Guideline Instability Electrical and Mechanical review ...
Страница 165: ...Troubleshooting 155 A1 Checking for Unstable Components in a Weighing System ...
Страница 166: ...156 Chapter 9 B Guidelines for Electrical Mechanical or Configuration Issues ...
Страница 167: ...Troubleshooting 157 B1 Guidelines to Verify Electrical Installation ...
Страница 168: ...158 Chapter 9 B2 Guidelines to Verify Mechanical Installation ...
Страница 169: ...Troubleshooting 159 B3 Guidelines to Verify Configuration Filter Settings to Improve Stability ...
Страница 170: ...160 Chapter 9 C Integrated Technician and Stability Test Overview ...
Страница 171: ...Troubleshooting 161 E Testing for Non Return to Zero System with IT Summing Card ...
Страница 173: ...Troubleshooting 163 G Calibration Errors When Performing the Calibration ...
Страница 174: ...164 Chapter 9 H Mechanical Installation ...
Страница 175: ...Troubleshooting 165 J Electrical Inspection ...
Страница 176: ...166 Chapter 9 K Installation Check Points ...
Страница 177: ...Troubleshooting 167 M Weight Processor s Front Display is Blank or Locked ...
Страница 178: ...168 Chapter 9 N Analog Out HI 6510 Only ...
Страница 194: ...184 Appendix A Figure D1 Assembling the Swivel Mount Enclosure on the instrument pedestal ...
Страница 195: ...Installation of Swivel Mount Enclosures 185 Figure D2 Assembling the Swivel Mount Enclosure on a Table ...
Страница 196: ...186 Appendix A Dimensions ...
Страница 200: ...4 HI 6500 Series USER GUIDE Index ...
Страница 201: ......