Instrument Configuration
103
•
•
•
•
•
•
Tare Command
When the Tare command is selected the Net weight value is set equal to 0 (zero), and the
displayed weight goes to 0.00 if the Net weight is being displayed. If the Gross weight is
being displayed the displayed value does not change.
On the
Web page
, left click on the Tare button located in the Operations page.
NOTE
The Tare operation will fail if the scale is in motion.
On the
HI 6500 series instrument display
, select the calibration menu and use the UP or
DOWN button to select the Tare Operations submenu, then select the Tare menu item and
press the ENTER button to execute this command.
Zero Operations Commands and Parameters
Zero Tolerance Parameter
The value you enter for Zero Tolerance sets the weight unit limit from zero the instrument
will accept as gross zero during the zeroing function (when you push the Zero button).
NOTE
The amount of weight zeroed off is cumulative. The zero command will fail if the current
gross weight plus any previously zeroed amount exceeds the zero tolerance.
RANGE: .000001 - 999999 . (default 10.0)
On the
Web page
, select the Operations menu and enter the value in the text box provided.
Our example, for illustration only, sets the Zero Tolerance to 10.0
On the
HI 6500 series instrument display
, select the Operations menu and use the UP or
DOWN button to select the Zero Operations submenu, then select the Zero Tolerance menu
item. The current value will be displayed, press the ENTER or the RIGHT button if this
value needs to be modified.
Zero Amount Parameter
The Zero Amount parameter is READ ONLY and displays the amount that has been
"zeroed" from the scale.
Содержание HI 6500 Series
Страница 10: ...viii Contents ...
Страница 20: ...10 Chapter 2 ...
Страница 72: ...62 Chapter 4 ...
Страница 96: ...86 Chapter 5 Profibus DP Commands and Parameters ...
Страница 97: ...Instrument Configuration 87 Serial Option Termination ...
Страница 126: ...116 Chapter 5 ...
Страница 156: ...146 Chapter 8 ...
Страница 163: ...Troubleshooting 153 General Troubleshooting Flow Chart Index ...
Страница 164: ...154 Chapter 9 A Guideline Instability Electrical and Mechanical review ...
Страница 165: ...Troubleshooting 155 A1 Checking for Unstable Components in a Weighing System ...
Страница 166: ...156 Chapter 9 B Guidelines for Electrical Mechanical or Configuration Issues ...
Страница 167: ...Troubleshooting 157 B1 Guidelines to Verify Electrical Installation ...
Страница 168: ...158 Chapter 9 B2 Guidelines to Verify Mechanical Installation ...
Страница 169: ...Troubleshooting 159 B3 Guidelines to Verify Configuration Filter Settings to Improve Stability ...
Страница 170: ...160 Chapter 9 C Integrated Technician and Stability Test Overview ...
Страница 171: ...Troubleshooting 161 E Testing for Non Return to Zero System with IT Summing Card ...
Страница 173: ...Troubleshooting 163 G Calibration Errors When Performing the Calibration ...
Страница 174: ...164 Chapter 9 H Mechanical Installation ...
Страница 175: ...Troubleshooting 165 J Electrical Inspection ...
Страница 176: ...166 Chapter 9 K Installation Check Points ...
Страница 177: ...Troubleshooting 167 M Weight Processor s Front Display is Blank or Locked ...
Страница 178: ...168 Chapter 9 N Analog Out HI 6510 Only ...
Страница 194: ...184 Appendix A Figure D1 Assembling the Swivel Mount Enclosure on the instrument pedestal ...
Страница 195: ...Installation of Swivel Mount Enclosures 185 Figure D2 Assembling the Swivel Mount Enclosure on a Table ...
Страница 196: ...186 Appendix A Dimensions ...
Страница 200: ...4 HI 6500 Series USER GUIDE Index ...
Страница 201: ......