Troubleshooting
151
•
•
•
•
•
•
mV
DC voltage signals are between 0-25 millivolts. Overloads and negative millivolt
readings are not shown as actual readings but 25.5 mVfor over voltage and 0.0 for negative
voltage. You will need to use a multimeter with a 200 or 300mVDC range to view the out
of range voltages.
RTZ (Return to Zero) Test
This check is to see if the scale returns to a zero reading condition when it is empty. Run
this test only when the scale is empty. When the scale is calibrated at the
ZERO
or
Reference zero, there is a period during which the
WEIGHT
readings are collected. After
a valid calibration has been completed, the test compares the original weight reading with
collected and current readings. If the difference is more than the combined Motion and the
Zero Tolerance parameters the test will fail.
IT Test
If your system has an Integrated Technician Summing Junction box, the IT test can identify
individual load cell problems up to a maximum of four load cell selections. The dwell
displays are live readings and only available if you have an IT summing junction card, with
a maximum of one JBox and 4 channels. From the front display, select Jbox1.
From the Diagnostics menu, select Integrated Technician.
Sensor Number
Indicates which sensor is under test. Using the up or down arrow selects
the target sensor to be tested.Then select the number of load cells to test. and press Do IT
Test. The Dwell test for pass or fail take 30 seconds to settle. Until it settles, it will always
show “fail” when the test is started. If the browser times out during the test, refresh your
browser until the results appear.
Содержание HI 6500 Series
Страница 10: ...viii Contents ...
Страница 20: ...10 Chapter 2 ...
Страница 72: ...62 Chapter 4 ...
Страница 96: ...86 Chapter 5 Profibus DP Commands and Parameters ...
Страница 97: ...Instrument Configuration 87 Serial Option Termination ...
Страница 126: ...116 Chapter 5 ...
Страница 156: ...146 Chapter 8 ...
Страница 163: ...Troubleshooting 153 General Troubleshooting Flow Chart Index ...
Страница 164: ...154 Chapter 9 A Guideline Instability Electrical and Mechanical review ...
Страница 165: ...Troubleshooting 155 A1 Checking for Unstable Components in a Weighing System ...
Страница 166: ...156 Chapter 9 B Guidelines for Electrical Mechanical or Configuration Issues ...
Страница 167: ...Troubleshooting 157 B1 Guidelines to Verify Electrical Installation ...
Страница 168: ...158 Chapter 9 B2 Guidelines to Verify Mechanical Installation ...
Страница 169: ...Troubleshooting 159 B3 Guidelines to Verify Configuration Filter Settings to Improve Stability ...
Страница 170: ...160 Chapter 9 C Integrated Technician and Stability Test Overview ...
Страница 171: ...Troubleshooting 161 E Testing for Non Return to Zero System with IT Summing Card ...
Страница 173: ...Troubleshooting 163 G Calibration Errors When Performing the Calibration ...
Страница 174: ...164 Chapter 9 H Mechanical Installation ...
Страница 175: ...Troubleshooting 165 J Electrical Inspection ...
Страница 176: ...166 Chapter 9 K Installation Check Points ...
Страница 177: ...Troubleshooting 167 M Weight Processor s Front Display is Blank or Locked ...
Страница 178: ...168 Chapter 9 N Analog Out HI 6510 Only ...
Страница 194: ...184 Appendix A Figure D1 Assembling the Swivel Mount Enclosure on the instrument pedestal ...
Страница 195: ...Installation of Swivel Mount Enclosures 185 Figure D2 Assembling the Swivel Mount Enclosure on a Table ...
Страница 196: ...186 Appendix A Dimensions ...
Страница 200: ...4 HI 6500 Series USER GUIDE Index ...
Страница 201: ......