70
Chapter
5
•
•
•
•
•
•
Calibration Menu
•
C2 eCAL
o Sensitivity
o Gravity
o Ref Wt
o Do C2 Calibration
o Cal Tol
o Num Dev
•
Hard Cal
o Cal Tol
o Sensitivity
o Cal Lo Wt
o Do Cal Lo
o Span Wt
o Do Cal Hi
•
Cal Date
•
Multi-Point Calibration
C2 (eCal) Commands and Parameters
Cal Tolerance Parameter
Sometimes, the contents of the vessel you are weighing are in motion. Cal Tolerance allows
you to set a value that determines the amount of motion that the system can tolerate and still
calibrate. In other words, the value you enter for Calibration Tolerance sets the amount of
deviation to allow during the calibration process. This value must be greater than or equal
to the base motion value and/or the Graduation Sizes.
The base motion value can be calculated by using the following formula:
Base Motion value = (Total Load Cell Capacity x 0.0003)
RANGE: .000001 - 999999 (default 10)
Another way to determine what this value should be is to multiply the fluctuation from peak
to peak and multiply this amount by two
Содержание HI 6500 Series
Страница 10: ...viii Contents ...
Страница 20: ...10 Chapter 2 ...
Страница 72: ...62 Chapter 4 ...
Страница 96: ...86 Chapter 5 Profibus DP Commands and Parameters ...
Страница 97: ...Instrument Configuration 87 Serial Option Termination ...
Страница 126: ...116 Chapter 5 ...
Страница 156: ...146 Chapter 8 ...
Страница 163: ...Troubleshooting 153 General Troubleshooting Flow Chart Index ...
Страница 164: ...154 Chapter 9 A Guideline Instability Electrical and Mechanical review ...
Страница 165: ...Troubleshooting 155 A1 Checking for Unstable Components in a Weighing System ...
Страница 166: ...156 Chapter 9 B Guidelines for Electrical Mechanical or Configuration Issues ...
Страница 167: ...Troubleshooting 157 B1 Guidelines to Verify Electrical Installation ...
Страница 168: ...158 Chapter 9 B2 Guidelines to Verify Mechanical Installation ...
Страница 169: ...Troubleshooting 159 B3 Guidelines to Verify Configuration Filter Settings to Improve Stability ...
Страница 170: ...160 Chapter 9 C Integrated Technician and Stability Test Overview ...
Страница 171: ...Troubleshooting 161 E Testing for Non Return to Zero System with IT Summing Card ...
Страница 173: ...Troubleshooting 163 G Calibration Errors When Performing the Calibration ...
Страница 174: ...164 Chapter 9 H Mechanical Installation ...
Страница 175: ...Troubleshooting 165 J Electrical Inspection ...
Страница 176: ...166 Chapter 9 K Installation Check Points ...
Страница 177: ...Troubleshooting 167 M Weight Processor s Front Display is Blank or Locked ...
Страница 178: ...168 Chapter 9 N Analog Out HI 6510 Only ...
Страница 194: ...184 Appendix A Figure D1 Assembling the Swivel Mount Enclosure on the instrument pedestal ...
Страница 195: ...Installation of Swivel Mount Enclosures 185 Figure D2 Assembling the Swivel Mount Enclosure on a Table ...
Страница 196: ...186 Appendix A Dimensions ...
Страница 200: ...4 HI 6500 Series USER GUIDE Index ...
Страница 201: ......