Network Configuration
59
•
•
•
•
•
•
Printer (over RS-232)
The HI 6500 series can transmit Gross, Net, Count, Tare values to a serial printer over RS-
232. The printer Baud rate is user selectable at 600, 1200, 2400, 4800, 9600, or 19200
(default is 9600).
The RS-232 printer pinout is:
A sample wiring connection from the HI 6500 series
to the printer port is shown above.
The printer parameters can be configured through the web interface or the display panel.
Step 1.
Set the baud rate for the RS232 printer, default is 9600.
Step 2.
Set the parity bit, default is EVEN
Step 3.
Select what values to transmit over the serial interface to the printer using the
printer mode, default is disabled.
a. Gross Weight
b. Net Weight
c. Tare Weight
d. Unit Count
e. All – Gross, Net, Tare, and Count (if Count enabled) values
f. OFF, Printer disabled.
USB Memory Stick
The HI 6500 series provides an interface to an external USB memory stick that allows
HI 6500 parameters to be saved, restored. or copied to another instrument.
NOTE
The HI 6500 series does not include a USB memory stick port when used in a Swivel Mount
Enclosure. See Appendix A for further information.
The USB memory stick commands can be activated through the web interface or the
display panel.
To access the USB Menu, from the front keypad, select Configuration > Enter,
Communications > Enter, USB >Enter, UP or DOWN arrow to Save or Restore > Enter.
On the
Web page
: Configuration; communications; USB; save/restore.
GND
DNC
TR+
TR-
PrtRX
PrtTX
Signal
Ground
DO Not
Connect
RS485 Pins Not Used Printer RX Printer TX
Содержание HI 6500 Series
Страница 10: ...viii Contents ...
Страница 20: ...10 Chapter 2 ...
Страница 72: ...62 Chapter 4 ...
Страница 96: ...86 Chapter 5 Profibus DP Commands and Parameters ...
Страница 97: ...Instrument Configuration 87 Serial Option Termination ...
Страница 126: ...116 Chapter 5 ...
Страница 156: ...146 Chapter 8 ...
Страница 163: ...Troubleshooting 153 General Troubleshooting Flow Chart Index ...
Страница 164: ...154 Chapter 9 A Guideline Instability Electrical and Mechanical review ...
Страница 165: ...Troubleshooting 155 A1 Checking for Unstable Components in a Weighing System ...
Страница 166: ...156 Chapter 9 B Guidelines for Electrical Mechanical or Configuration Issues ...
Страница 167: ...Troubleshooting 157 B1 Guidelines to Verify Electrical Installation ...
Страница 168: ...158 Chapter 9 B2 Guidelines to Verify Mechanical Installation ...
Страница 169: ...Troubleshooting 159 B3 Guidelines to Verify Configuration Filter Settings to Improve Stability ...
Страница 170: ...160 Chapter 9 C Integrated Technician and Stability Test Overview ...
Страница 171: ...Troubleshooting 161 E Testing for Non Return to Zero System with IT Summing Card ...
Страница 173: ...Troubleshooting 163 G Calibration Errors When Performing the Calibration ...
Страница 174: ...164 Chapter 9 H Mechanical Installation ...
Страница 175: ...Troubleshooting 165 J Electrical Inspection ...
Страница 176: ...166 Chapter 9 K Installation Check Points ...
Страница 177: ...Troubleshooting 167 M Weight Processor s Front Display is Blank or Locked ...
Страница 178: ...168 Chapter 9 N Analog Out HI 6510 Only ...
Страница 194: ...184 Appendix A Figure D1 Assembling the Swivel Mount Enclosure on the instrument pedestal ...
Страница 195: ...Installation of Swivel Mount Enclosures 185 Figure D2 Assembling the Swivel Mount Enclosure on a Table ...
Страница 196: ...186 Appendix A Dimensions ...
Страница 200: ...4 HI 6500 Series USER GUIDE Index ...
Страница 201: ......