76
Chapter
5
•
•
•
•
•
•
Span Weight Parameter
The Span Weight value is the weight of the object that is being placed on the scale to
set the “High” calibration point with respect to the “Low” calibration point set using
the Cal Lo Weight parameter.
On the
Web page
, select the calibration menu and left click inside the Span Weight
text field and enter the load cell sensitivity value.
On the
HI 6500 series instrument display
, select the calibration menu and use the UP or
DOWN button to select the Hard Cal submenu, then select the Span weight menu item. The
current Span Weight value will be displayed, press the ENTER or the RIGHT button if this
value needs to be modified.
Cal Date Parameter
The Cal Date parameter is used to record the date when the last calibration was completed
on the HI 6500 series instrument. The format is
mm/dd/yyyy
On the
Web page
, select the calibration menu and left click inside the Cal Date text field
and enter the date the calibration was completed.
On the
HI 6500 series instrument display
, select the calibration menu and use the UP or
DOWN button to select the Cal Date submenu. The last Cal Date value will be displayed,
press the ENTER or the RIGHT button if this date needs to be modified.
Multi-Point Calibration
For details on Multi-Point Calibration, see page 123.
Содержание HI 6500 Series
Страница 10: ...viii Contents ...
Страница 20: ...10 Chapter 2 ...
Страница 72: ...62 Chapter 4 ...
Страница 96: ...86 Chapter 5 Profibus DP Commands and Parameters ...
Страница 97: ...Instrument Configuration 87 Serial Option Termination ...
Страница 126: ...116 Chapter 5 ...
Страница 156: ...146 Chapter 8 ...
Страница 163: ...Troubleshooting 153 General Troubleshooting Flow Chart Index ...
Страница 164: ...154 Chapter 9 A Guideline Instability Electrical and Mechanical review ...
Страница 165: ...Troubleshooting 155 A1 Checking for Unstable Components in a Weighing System ...
Страница 166: ...156 Chapter 9 B Guidelines for Electrical Mechanical or Configuration Issues ...
Страница 167: ...Troubleshooting 157 B1 Guidelines to Verify Electrical Installation ...
Страница 168: ...158 Chapter 9 B2 Guidelines to Verify Mechanical Installation ...
Страница 169: ...Troubleshooting 159 B3 Guidelines to Verify Configuration Filter Settings to Improve Stability ...
Страница 170: ...160 Chapter 9 C Integrated Technician and Stability Test Overview ...
Страница 171: ...Troubleshooting 161 E Testing for Non Return to Zero System with IT Summing Card ...
Страница 173: ...Troubleshooting 163 G Calibration Errors When Performing the Calibration ...
Страница 174: ...164 Chapter 9 H Mechanical Installation ...
Страница 175: ...Troubleshooting 165 J Electrical Inspection ...
Страница 176: ...166 Chapter 9 K Installation Check Points ...
Страница 177: ...Troubleshooting 167 M Weight Processor s Front Display is Blank or Locked ...
Страница 178: ...168 Chapter 9 N Analog Out HI 6510 Only ...
Страница 194: ...184 Appendix A Figure D1 Assembling the Swivel Mount Enclosure on the instrument pedestal ...
Страница 195: ...Installation of Swivel Mount Enclosures 185 Figure D2 Assembling the Swivel Mount Enclosure on a Table ...
Страница 196: ...186 Appendix A Dimensions ...
Страница 200: ...4 HI 6500 Series USER GUIDE Index ...
Страница 201: ......