4
GMC-I Messtechnik GmbH
Overview of the Test Instrument’s Scope of Functions
1
10/25 A R
PE
measurements are only possible with line voltages of 115/
230 V and line frequencies of 50/60 Hz.
2
Voltage measuring inputs
3
Connection for 2
nd
test probe P2 for 2-pole measurement
4
Measurement of time to trip isn’t possible in IT systems.
5
No checking for reversed polarity takes place when the EL1 adapter is
used.
6
Connection type not available with feature G02
7
For ME equipment
(m
edical
e
lectrical equipment) with own power supply
Key
APP
= application part
Alternative = alternative measurement (equivalent leakage
current measurement)
Differential = differential current measurement
Direct
= direct measurement
LN(TS)
= short-circuited conductors L and N at test socket
P1
= measurement with test probe P1
P1-P2
= 2-pole measurement with test probes P1 and P2
PE-P1
= measurement between PE and test probe P1
PE(TS)
= protective conductor at the test socket
PE(mains) = protective conductor at the mains connection
Scope of Delivery
Standard Version (country-specific)
1
Test instrument
SECULIFE ST PRO
1
Mains power cable
1
Test probe, 2 m, not coiled
1
USB cable, USB A to USB B, 1.0 m long
1
Plug-on alligator clip
1
KS17-ONE cable set
1
Calibration certificate
1
Condensed operating instructions
1
Comprehensive operating instructions available on the Inter-
net
1
Card with registration key for
IZYTRONIQ
BUSINESS Starter soft-
ware (for download from the web)
Sw
it
ch
Po
si
ti
on
Measuring Functions
Test Current/Voltage
Measurement
Type,
Connection Type
Single measurements,
rotary switch level: green
R
PE
R
PE
Protective conductor resistance
PE(TS) - P1 passive
PE(TS) - P1 (TS on)
PE(mains) - P1
6
PE(mains) - P1 clamp
2, 6
P1–P2
3
I
P
200 mA test current
10 A test current
1
(feature G01)
25 A test current
1
(feature G02)
R
INS
R
INS
Insulation resistance (
PC I/PC II)
LN(TS) - PE(TS)
LN(TS) - P1
P1–P2
3
PE(mains) - P1
PE(TS) - P1
LN(TS) - P1//PE(TS)
LN(TS) - APP
PE(mains) - APP
PE(TS) - APP
P1//PE(TS) - APP
P2 - APP
U
INS
Test voltage
I
PE
I
PE
Protective conductor current, RMS
Direct
Differential
Alternative
AT3-Adapter
2
Clamp
2
I
PE
~
AC component
I
PE=
DC component
U
LPE
Test voltage
U
Gen
Reference voltage (alternative)
I
T
I
T
Touch current, TRMS
Direct P1
Differential P1
Alternative P1
Perm. conn. P1
Alternative P1-P2
I
T
~
AC component
I
T=
DC component
U
LPE
Test voltage
U
Gen
Reference voltage (alternative)
I
E
I
E
Device leakage current, TRMS
Direct
Differential
Alternative
AT3-Adapter
2
Clamp
2
I
E
~
AC component
I
E=
DC component
U
LPE
Test voltage
U
Gen
Reference voltage (alternative)
I
A
I
A
Leakage current from the applied part TRMS
Direct P1
Direct APP
Alternative P1
Alternative APP
Perm. conn. P1
Perm. conn. APP
APP - P2
7
U
LPE
Test voltage
U
Gen
Voltage at application part
I
P
I
P
Patient leakage current, TRMS
Direct P1
Direct APP
Perm. conn. P1
Perm. conn. APP
I
P
~
AC component
I
P=
DC component
U
LPE
Test voltage
I
PA
I
PA
Patient auxiliary current, TRMS
Direct APP
Perm. conn. APP
I
PA
~
AC component
I
PA=
DC component
U
LPE
Test voltage
U
U
Probe voltage, RMS
PE–P1
PE–P1 (TS on *)
* Polarity parameter
U
~
Alternating voltage component
U
=
Direct voltage component
U
Measuring voltage, RMS
2
V – COM
V – COM (TS on)
U
~
Alternating voltage component
2
U
=
Direct voltage component
2
P
Function test at the test socket
Polarity parameter
I
Current between L and N
U
Voltage between L and N
f
Frequency
P
Active power
S
Apparent power
PF
Power factor
Probe measuring functions
EL1
Function test for extension cords
with adapter:
continuity, short-circuit, polarity (wire reversal
5
)
EL1 adapter
AT3-IIIE adapter
VL2E adapter
EXTRA
Reserved for expansion within the framework of software updates
t
A
4
PRCD time to trip for 10/30 mA PRCDs
°C
Temperature measurement
2
with Pt100/Pt1000 V – COM
IZ
Current clamp measurement
2
with current clamp sensor V – COM
Sw
it
ch
Po
si
ti
on
Standard
Measurement Type, Connection Type
Automatic test sequences,
rotary switch level: orange
Preconfigured (freely adjustable) test sequences – default settings
A1
IEC 62353
Passive, test socket, 1 group BF APPs A-K, PCI
A2
IEC 62353
Passive, test socket, 1 group BF APPs A-K, PCII
A3
IEC 62353
Passive, tests on, 1 group BF APPs A-K, PCI + II
A4
IEC 62353
Active, automatic detection, 1 group BF APPs A-K, PCI
A5
IEC 62353
Active, automatic detection, 1 group BF APPs A-K, PCII
A6
IEC 62353
Active, automatic detection, 1 group BF APPs A-K, PCI + II
With feature D02
A7
VDE 0701-0702
ÖVE E 8701
SNR 462638
Active, automatic detection of the DUT connection, PCI + II
A8
VDE 0701-0702
ÖVE E 8701
SNR 462638
Active, automatic detection of the DUT connection, PCI + II
A9
VDE 0701-0702
ÖVE E 8701-EDV
SNR 462638-EDV
Active, automatic detection of the DUT connection, PCI + II
With feature KA01
A7
IEC 60601
3
rd
edition
Active, automatic detection, 1 group BF APPs A-K, PCI
A8
IEC 60601
3
rd
edition
Active, automatic detection, 1 group BF APPs A-K, PCII
A9
IEC 60601
3
rd
edition
Active, automatic detection, 1 group BF APPs A-E, PCI + II
Active, automatic detection, 1 group CF APPs F-K, PCI + II
Содержание SECULIFE ST PRO