Alternative Measuring Method (equivalent patient leakage
current)
–
Measurement type: alternative P1
– DUT mains plug (PC1) connected to test socket
– Probe to P1 terminal
Schematic Diagram
After activating test voltage, leakage current from the application
part is measured between short-circuited conductors L-N-PE
(DUT mains plug) and the short-circuited terminals of the applied
parts (test probe P1).
Wiring Diagram
Alternative Measuring Method (equivalent patient leakage
current)
–
Measurement type: alternative APP
– DUT mains plug (PC1) connected to test socket
– APPs to APP sockets
Schematic Diagram
After activating test voltage, leakage current from the applied part
is measured between short-circuited conductors L-N-PE (DUT
mains plug) and the applied parts connected to the APP sockets
at the test instrument.
Wiring Diagram
Direct Measuring Method
–
Measurement type: permanent connection P1
– Permanent connection
– Probe to P1 terminal
Schematic Diagram
Leakage current from the applied part is measured between the
short-circuited terminals of the applied parts (test probe P1) and
PE at the mains connection.
Wiring Diagram
Direct Measuring Method
–
Measurement type: permanent connection APP
– Permanent connection
– APPs to APP sockets
Schematic Diagram
Leakage current from the applied part is measured between the
applied parts connected to the APP sockets at the test instru-
ment and PE at the mains connection.
Accessible conductive parts which are not connected to the
housing must be contacted with test probe P1.
Wiring Diagram