Direct Measuring Method
–
Measurement type APP - P2
– DUT mains plug (PC1) connected to test socket
– Probe to P2 terminal
– APPs to APP sockets
Schematic Diagram
Leakage current from the applied part is measured between the
applied parts connected to the APP sockets at the test instru-
ment and exposed, conductive parts (test probe P2), which are
not connected to the housing.
Wiring Diagram
Setting Measuring Parameters for IA
1
For ME equipment (
m
edical
e
lectrical equipment)
with own power supply
2
This measurement only detects the portion of the leakage current which is gen-
erated due voltage coupling at the DUT’s measuring input.
Measuring
Parameter
Meaning
Measurement Type
Suitable for
DUT Connection via
Direct P1
Direct measurement of leakage
current from the applied part via
probe P1
Test socket, AT3 adapter
(AT3-IIIE, AT3-IIS, AT3-IIS32),
AT16DI/AT32DI
Direct APP
Direct measurement of leakage
current from the applied part via
selected APP socket(s)
Test socket
Alternative P1
Equivalent measurement of leak-
age current from the applied part
via probe P1
Test socket
Alternative APP
Equivalent measurement of leak-
age current from the applied part
via selected APP socket(s)
Test socket
Perm. conn. P1
Direct measurement of leakage
current from the applied part via
probe P1
Permanent connection
Perm. conn. APP
Direct measurement of leakage
current from the applied part via
selected APP socket(s)
Permanent connection
APP – P2
1
Direct measurement of leakage
current from the applied part be-
tween selected APP socket(s)
and probe P2
APP-P2
APP – applied parts
Individual selection: A / B / C / D / E / F / G / H / I / K, respectively via on/off
Phase angle – for direct and permanent connection measurement types
only
0° or 180°
Selectable phasing for the internal generator relative to mains phas-
ing
Polarity
With direct only
L/N or N/L
Selection of polarity for mains voltage to the test socket