78
GMC-I Messtechnik GmbH
10.4
Selecting and Configuring a Test Sequence
Sample: Initial Page of a Test Sequence – Tree View
Sample: Initial Page of a Test Sequence – Detail View and Applied
Part
* SETUP switch position:
Setup Menu 1/3 > Auto Measurements > 2/2 > Initial Window Style:
Tree
or
Detail View
Test instruments with feature E01 (touchscreen)
The display can be switched back and forth between the “tree
view” and the “detail view” (see above) via “Touch Click”, i.e. by
briefly tapping within the bottom frame.
Classification Parameter – Auto-Detection
If the settings for certain classification parameters are automati-
cally detected by the test instrument, this is indicated in each case by
an orange frame. Descriptions of these parameters are listed in the
following tables relative to the respective switch positions.
Automatic Detection Active for Protection Class
When connecting or disconnecting a DUT, the pro-
tection class can be changed without prior authori-
zation.
Automatic Detection Inactive for Protection Class
The test instrument retains the selected protection
class setting when a DUT is connected or discon-
nected.
Conveniently Changing Classification Parameters
(feature E01, “touchscreen”)
➭
The corresponding selection menu appears after touch click-
ing (briefly tapping) the respective classification parameters
window.
➭
The display is automatically returned to the start menu after
selecting the desired parameter.
Classification Parameter – VDE 0701-0702
1
These parameters must be entered manually if they’re not automatically
detected, or if they’re detected incorrectly.
2
The limit value for protective conductor resistance is determined on the
basis of length and cross-section. Data remain in memory until a new
entry is made.
A1
Standard / Test Sequence
Connection Type
Protection Class
Cl. Parameter
Measurement Type
Status Line
Sequence Parameters
Tree View *
A1
Connection Type
Protection Class
Cl. Parameter
Measurement Type
Status Line
Sequence Parameters
Applied Parts
Detail View *
Standard / Test Sequence
Parameter
Setting Options / Meaning
1/2
Standard
VDE 0701-0702
VDE 0701-0702-EDV, see following table
IEC 60601, 2
nd
edition, see table below
EC 60601, 3
rd
edition, see table below
EC 60601, 3
rd
edition GPA, see table below
IEC 62353, see table below
VDE 0701-0702-VLTG, see table below
VDE 0701-0702-PRCD, see table below
IEC 60974-4, see table below
Protection class
1, 2
PCI, PCII, PCI+II, PCI+III, PCII+III, PCI+II+III
Connection type
1, 2
Test socket
Permanent connection
Adapter: AT16/32-DI adapter
Adapter: VL2E
Adapter: AT3-Adapter (feature I01)
Permanent connection: P1+P2 (only with feature H01)
2/2
Measurement type (MT)
1
Passive
Active
Auto-detection
Connection & PC & MT
Connection & PC
Connection & MT
Connection only
PC & measurement type
Protection class only (PC)
Measurement type only (MT)
Disabled: No auto-detection:
All classification parameters such as connection, protection
class and measurement type must be entered manually.