60
GMC-I Messtechnik GmbH
– Special Case, Single Fault: Voltage at Applied Parts
Schematic Diagram
The fault condition
Voltage at Applied Parts
between APP and PE is
shown here.
Wiring Diagram
Setting Measuring Parameters for IP
Note
The currently selected
single fault
and the
test condition
are
displayed on the initial page. In order to change them,
they have to be selected and
set via the measuring
parameters menu
.
The measurements have to be conducted under all fault condi-
tions. Selection is made in the measuring parameters menu under
Single Fault
.
Measuring
Parameter
Meaning
Measurement Type
Suitable for
DUT Connection via
Direct P1
Direct measuring method (via
test socket) with test probe P1
Test socket
Direct APP
Direct measuring method (via
test socket) via selected APP
socket(s).
Test socket
Perm. conn. P1
Measurement at permanently
installed DUT with test probe P1
Permanent connection
Perm. conn. APP
Measurement at permanently
installed DUT via selected APP
socket(s).
Permanent connection
Test conditions – for measurement type with APP only
None
APP > PE: Connect applied part to PE –
all
APP sockets are connected to PE potential
when this option is activated.
Housing > PE: Connect housing to PE
APP/Housing > PE: Connect housing and applied part to PE
APP – applied parts
Individual selection: A / B / C / D / E / F / G / H / I / K, respectively via on/off
Single fault (SFC)
Normal state / N interrupted / PE interrupted / voltage at APPs
Polarity
Only with direct measurement
type
L/N or N/L
Selection of polarity for mains voltage to the test socket